Alere Triage CardioProfiler Panel PN 97100CP Used for the quantitative determination of creatine kinase MB, myoglobin, troponin I and B-type natriuretic peptide in EDTA anticoagulated whole blood and plasma specimens.
Product Description
Alere Triage CardioProfiler Panel PN 97100CP Used for the quantitative determination of creatine kinase MB, myoglobin, troponin I and B-type natriuretic peptide in EDTA anticoagulated whole blood and plasma specimens.
Reason for Recall
Certain lots of the affected products may have significantly decreased precision relative to the package insert, which could result in an increased frequency of false positive or false negative results. These false positive and false negative results are unpredictable within lots and may not be detected by Quality Control testing.
Details
- Recalling Firm
- Alere San Diego, Inc.
- Units Affected
- 16,962 kits
- Distribution
- Nationwide Distribution
- Location
- San Diego, CA
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11