VITROS Chemistry Products CKMB Slides, Creatine Kinase MB, Product Code 8001133-90 slides, UDI Number 10758750004201 - Product Usage: The VITROS Chemistry Product Slides are used to measure creatine kinase MB (CK-MB) activity in serum which used to be ordered to detect Myocardial Infarction.
Reported: May 20, 2020 Initiated: April 1, 2020 #Z-1931-2020
Product Description
VITROS Chemistry Products CKMB Slides, Creatine Kinase MB, Product Code 8001133-90 slides, UDI Number 10758750004201 - Product Usage: The VITROS Chemistry Product Slides are used to measure creatine kinase MB (CK-MB) activity in serum which used to be ordered to detect Myocardial Infarction.
Reason for Recall
The Spanish and Swedish language versions of the IFU state that the assay can be tested with Plasma which is false and could result in biased results.
Details
- Recalling Firm
- Ortho-Clinical Diagnostics
- Units Affected
- N/A
- Distribution
- US Nationwide distribution.
- Location
- Rochester, NY
Frequently Asked Questions
What product was recalled? ▼
VITROS Chemistry Products CKMB Slides, Creatine Kinase MB, Product Code 8001133-90 slides, UDI Number 10758750004201 - Product Usage: The VITROS Chemistry Product Slides are used to measure creatine kinase MB (CK-MB) activity in serum which used to be ordered to detect Myocardial Infarction.. Recalled by Ortho-Clinical Diagnostics. Units affected: N/A.
Why was this product recalled? ▼
The Spanish and Swedish language versions of the IFU state that the assay can be tested with Plasma which is false and could result in biased results.
Which agency issued this recall? ▼
This recall was issued by the FDA Devices on May 20, 2020. Severity: Moderate. Recall number: Z-1931-2020.
Related Recalls
FDA Devices Moderate
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
FDA Devices Moderate
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
FDA Devices Moderate
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11