FDA Devices Moderate Class II Terminated

SOMATOM Emotion 16; produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angels or spiral planes taken at different angles.

Reported: July 8, 2015 Initiated: June 4, 2015 #Z-1933-2015

Product Description

SOMATOM Emotion 16; produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angels or spiral planes taken at different angles.

Reason for Recall

A potential safety issue involving two (2) broken screws and one (1) bowed, which is used to fix the Balance Weight of the DMS in the gantry, was found during a service call at the customer site.

Details

Units Affected
40 total
Distribution
Nationwide Distribution.
Location
Malvern, PA

Frequently Asked Questions

What product was recalled?
SOMATOM Emotion 16; produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angels or spiral planes taken at different angles.. Recalled by Siemens Medical Solutions USA, Inc. Units affected: 40 total.
Why was this product recalled?
A potential safety issue involving two (2) broken screws and one (1) bowed, which is used to fix the Balance Weight of the DMS in the gantry, was found during a service call at the customer site.
Which agency issued this recall?
This recall was issued by the FDA Devices on July 8, 2015. Severity: Moderate. Recall number: Z-1933-2015.