Severity
Moderate
FDA Devices recall · Reported June 6, 2018
There is a possibility that DDK2-2.4-30X Disposable Drill Kits, Lot Number 111664 208140649 contained 2.8mm drill bits from DDK2-2.8-30X Disposable Drill Kits, Lot Number 111745 2…
Ad-Tech Medical Instrument Corporation recalled Disposable Drill Kit which includes two drill bits, two drill stops, and two adjustment w… - a moderate-severity action.
Disposable Drill Kit which includes two drill bits, two drill stops, and two adjustment w… was recalled by Ad-Tech Medical Instrument Corporation in June 6, 2018. Reason: There is a possibility that DDK2-2.4-30X Disposable Drill Kits, Lot Number 111664 208140649 contained 2.8mm d…. Check the official notice for the remedy. Verify recall #Z-1938-2018 with the FDA Devices before acting.
The recall
Ad-Tech Medical Instrument Corporation issued this moderate-severity FDA Devices recall-There is a possibility that DDK2-2.4-30X Disposable Drill Kits, Lot Number 111664 208140649 contained 2.8mm d….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1938-2018 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1938-2018) was formally reported on June 6, 2018, with the manufacturer initiating the action on December 20, 2017. It is classified under Moderate severity (Class II), with a current status of Terminated. Ad-Tech Medical Instrument Corporation is listed as the recalling firm, operating out of Oak Creek, WI. Federal records list the affected scope as 30 devices.
The documented reason for this recall is: There is a possibility that DDK2-2.4-30X Disposable Drill Kits, Lot Number 111664 208140649 contained 2.8mm drill bits from DDK2-2.8-30X Disposable Drill Kits, Lot Number 111745 208140649 and vice versa. Distribution data in the federal record shows the product reached: Worldwide Distribution: US (nationwide) to states of: CA, DC, FL, MI, MA, NC, NY, WA, WI; and country of: Canada. Distributed internationally to the United Kingdom.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
30 devices
Related Recalls
6
3 from same agency
Disposable Drill Kit which includes two drill bits, two drill stops, and two adjustment wrenches, sterile. The drill bits are 30cm long and made of stainless steel. The drill kits are supplied in a sterile state and are single-use only. Used to drill cranial holes using a stereotactic frame.
There is a possibility that DDK2-2.4-30X Disposable Drill Kits, Lot Number 111664 208140649 contained 2.8mm drill bits from DDK2-2.8-30X Disposable Drill Kits, Lot Number 111745 208140649 and vice versa.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-1938-2018 |
| Date reported | June 6, 2018 |
| Date initiated | December 20, 2017 |
| Recalling firm | Ad-Tech Medical Instrument Corporation |
| Firm location | Oak Creek, WI |
| Affected scope | 30 devices |
| Distribution | Worldwide Distribution: US (nationwide) to states of: CA, DC, FL, MI, MA, NC, NY, WA, WI; and country of: Canada. Distributed internationally to the United Kingdom. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported June 6, 2018. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.