PlainRecalls
FDA Devices Moderate Class II Terminated

PAC Tray/5 Catheter Introducer Tray 8F, Model 497424, Rx Only, STERILE EO, UDI: 00886333214440 - Product Usage: The introducers are indicated for use in arterial and venous procedures requiring percutaneous introduction of intravascular devices. Curved tip Introducer allows for contralateral access to the iliac artery and selected vasculature. Case Label: PAC Tray/5 Catheter Introducer Tray 8F, QTY 10/Bx, Rx Only, STERILE EO

Reported: May 20, 2020 Initiated: April 8, 2020 #Z-1938-2020

Product Description

PAC Tray/5 Catheter Introducer Tray 8F, Model 497424, Rx Only, STERILE EO, UDI: 00886333214440 - Product Usage: The introducers are indicated for use in arterial and venous procedures requiring percutaneous introduction of intravascular devices. Curved tip Introducer allows for contralateral access to the iliac artery and selected vasculature. Case Label: PAC Tray/5 Catheter Introducer Tray 8F, QTY 10/Bx, Rx Only, STERILE EO

Reason for Recall

Arterial Line Kits include thicker curved reverse-cutting needle with a 0-silk suture, instead of the usual straight-cutting needle with a 3-0 silk suture. The thicker suture may be more difficult to tie. The curved needle requires the use of a needle driver, there is a risk of needle stick to the surgeon tries to do it with fingers instead.

Details

Units Affected
80
Distribution
US Nationwide distribution including the states of CA, IL, ND, SC, TX, VA. OUS: None
Location
Athens, TX

Frequently Asked Questions

What product was recalled?
PAC Tray/5 Catheter Introducer Tray 8F, Model 497424, Rx Only, STERILE EO, UDI: 00886333214440 - Product Usage: The introducers are indicated for use in arterial and venous procedures requiring percutaneous introduction of intravascular devices. Curved tip Introducer allows for contralateral access to the iliac artery and selected vasculature. Case Label: PAC Tray/5 Catheter Introducer Tray 8F, QTY 10/Bx, Rx Only, STERILE EO. Recalled by Argon Medical Devices, Inc. Units affected: 80.
Why was this product recalled?
Arterial Line Kits include thicker curved reverse-cutting needle with a 0-silk suture, instead of the usual straight-cutting needle with a 3-0 silk suture. The thicker suture may be more difficult to tie. The curved needle requires the use of a needle driver, there is a risk of needle stick to the surgeon tries to do it with fingers instead.
Which agency issued this recall?
This recall was issued by the FDA Devices on May 20, 2020. Severity: Moderate. Recall number: Z-1938-2020.