FDA Devices recall · Reported June 12, 2024
The Synchro Neuro Guidewires are a steerable guidewire family with a shapeable tip. The outside diameter of the Synchro 10 guidewire is 0.010in. The outside diameter of the Syncrho 14 guidewire is 0.014in. The guidewires are compatible with existing microcatheters used in common procedures such as those used in endovascular diagnosis and therapy of neurovascular disease. The distal portion of the guidewire tip is radiopaque. For lubricity, the distal portion of the device is coated with a hydrop
Stryker Neurovascular has observed an increased frequency of PTFE coating damage occurring on the Synchro Guidewires that may be caused by the practice of backloading the guidewir…
Stryker Neurovascular recalled The Synchro Neuro Guidewires are a steerable guidewire family with a shapeable tip. The o… - a moderate-severity action.
The Synchro Neuro Guidewires are a steerable guidewire family with a shapeable tip. The o… was recalled by Stryker Neurovascular in June 12, 2024. Reason: Stryker Neurovascular has observed an increased frequency of PTFE coating damage occurring on the Synchro Gui…. Check the official notice for the remedy. Verify recall #Z-1939-2024 with the FDA Devices before acting.
The recall
Stryker Neurovascular issued this moderate-severity FDA Devices recall-Stryker Neurovascular has observed an increased frequency of PTFE coating damage occurring on the Synchro Gui….
- Moderate
- severity level
- 105K units
- affected scope
- Class II
- classification
- June 12, 2024
- reported
Sourced from official FDA Devices enforcement records. Verify recall #Z-1939-2024 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
Recall Insight
This recall's affected scope (see the Recall Profile below) is large enough to require multi-state distribution tracking. Distribution scope (see Distribution below) directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Medical Devices recalls over time
Where this recall sits in its category - 41,021 medical devices recalls on record
Of 88,244 recalls with an agency hazard classification, 19,068 are high severity, 64,652 moderate, and 4,524 low. A further 16,179 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Product description
The Synchro Neuro Guidewires are a steerable guidewire family with a shapeable tip. The outside diameter of the Synchro 10 guidewire is 0.010in. The outside diameter of the Syncrho 14 guidewire is 0.014in. The guidewires are compatible with existing microcatheters used in common procedures such as those used in endovascular diagnosis and therapy of neurovascular disease. The distal portion of the guidewire tip is radiopaque. For lubricity, the distal portion of the device is coated with a hydrophilic polymer and the proximal portion of the guidewire is coated with polytetrafluoroethylene (PTFE). The torque device included with the guidewire attaches to the proximal end of the wire and functions as a steering guide. Rotation of this device facilitates guidewire placement into the appropriate vessel by precise directional manipulation of the guidewire tip. The introducer included with the guidewire is intended to aid insertion of the guidewire into the catheter hub and/or hemostasis valve.
Reason for recall
Stryker Neurovascular has observed an increased frequency of PTFE coating damage occurring on the Synchro Guidewires that may be caused by the practice of backloading the guidewire through the optional introducer accessory. This issue is limited to certain lots of the Synchro Guidewire that contain an older version of the introducer accessory. Due to variation in the manufacturing process of the supplier of the introducer accessory, certain lots of introducers have sharper than intended edges that can peel off the PTFE coating when physicians use a technique known as backloading. Users with impacted product in their inventory are cautioned not to use this backloading technique.
Recall Profile
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Ongoing |
| Recall number | Z-1939-2024 |
| Date reported | June 12, 2024 |
| Date initiated | April 18, 2024 |
| Recalling firm | Stryker Neurovascular |
| Firm location | Fremont, CA |
| Affected scope | 105,286 |
| Distribution | Worldwide - US Nationwide distribution including in the states of TX, NY, MD, CA, KY, WI, OR, NE, SC, MA, MO, CT, ID, PA, TN, FL, NC, AL, IA, OK, VA, IL, OH, NH, MN, ND, KS, NJ, WV, IN, GA, HI, AR, UT, LA, MI, WA, NV, CO, AZ, DC, VT, ME, D… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
- Check the recall number (Z-1939-2024) and product description against the item you own. Search the archive
- Confirm the current status with FDA Devices before acting, recall details can be updated.
- Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Where was the recalled product distributed? ▼
Should I stop using a recalled medication or medical device? ▼
Nationwide recalls with similar severity or category timing
Drug and device notices outside this firm and state, matched by severity timing and product category.
Similar severity · nearest report date
Same archive severity grade (Moderate), closest by report date among other firms and states.