Severity
Moderate
FDA Devices recall · Reported June 18, 2025
Investigation determined that one lot of the T2 Greater Trochanter Nail (GTN) was labeled as left version but contained a right nail and another lot was labeled as right version b…
Stryker GmbH recalled Femoral Nail, LEFT T2 GTN 8x300 mm. Catalog Number: 1850-0830S. - a moderate-severity action.
Femoral Nail, LEFT T2 GTN 8x300 mm. Catalog Number: 1850-0830S. was recalled by Stryker GmbH in June 18, 2025. Reason: Investigation determined that one lot of the T2 Greater Trochanter Nail (GTN) was labeled as left version but…. Check the official notice for the remedy. Verify recall #Z-1939-2025 with the FDA Devices before acting.
The recall
Stryker GmbH issued this moderate-severity FDA Devices recall-Investigation determined that one lot of the T2 Greater Trochanter Nail (GTN) was labeled as left version but….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1939-2025 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1939-2025) was formally reported on June 18, 2025, with the manufacturer initiating the action on May 22, 2025. It is classified under Moderate severity (Class II), with a current status of Ongoing. Stryker GmbH is listed as the recalling firm, operating out of Selzach. Federal records list the affected scope as 1.
The documented reason for this recall is: Investigation determined that one lot of the T2 Greater Trochanter Nail (GTN) was labeled as left version but contained a right nail and another lot was labeled as right version but contained a left nail. Distribution data in the federal record shows the product reached: Worldwide - US Nationwide distribution in the states of California and Missouri. The countries of Belgium, Japan, Netherlands.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
1
Related Recalls
6
3 from same agency
Femoral Nail, LEFT T2 GTN 8x300 mm. Catalog Number: 1850-0830S.
Investigation determined that one lot of the T2 Greater Trochanter Nail (GTN) was labeled as left version but contained a right nail and another lot was labeled as right version but contained a left nail.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Ongoing |
| Recall number | Z-1939-2025 |
| Date reported | June 18, 2025 |
| Date initiated | May 22, 2025 |
| Recalling firm | Stryker GmbH |
| Firm location | Selzach |
| Affected scope | 1 |
| Distribution | Worldwide - US Nationwide distribution in the states of California and Missouri. The countries of Belgium, Japan, Netherlands. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported June 18, 2025. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.