FDA Devices Moderate Class II Terminated

Syva¿ EMIT¿ 2000 Theophylline Product Usage: The Syva¿ Emit¿ 2000 Theophylline Assay is a homogeneous enzyme immunoassay intended for use in the quantitative analysis of theophylline in human serum or plasma.

Reported: June 6, 2018 Initiated: November 27, 2017 #Z-1946-2018

Product Description

Syva¿ EMIT¿ 2000 Theophylline Product Usage: The Syva¿ Emit¿ 2000 Theophylline Assay is a homogeneous enzyme immunoassay intended for use in the quantitative analysis of theophylline in human serum or plasma.

Reason for Recall

Lot J1 of the Syva¿ EMIT¿ 2000 Theophylline Assay, when run on a Beckman AU Clinical Chemistry System, may exhibit increased imprecision for commercially available controls and for patient samples.

Details

Units Affected
N/A
Distribution
US Nationwide Distribution
Location
Newark, DE

Frequently Asked Questions

What product was recalled?
Syva¿ EMIT¿ 2000 Theophylline Product Usage: The Syva¿ Emit¿ 2000 Theophylline Assay is a homogeneous enzyme immunoassay intended for use in the quantitative analysis of theophylline in human serum or plasma.. Recalled by Siemens Healthcare Diagnostics, Inc.. Units affected: N/A.
Why was this product recalled?
Lot J1 of the Syva¿ EMIT¿ 2000 Theophylline Assay, when run on a Beckman AU Clinical Chemistry System, may exhibit increased imprecision for commercially available controls and for patient samples.
Which agency issued this recall?
This recall was issued by the FDA Devices on June 6, 2018. Severity: Moderate. Recall number: Z-1946-2018.