FDA Devices recall · Reported August 21, 2013
CELLTRACKS ANALYZER II (Software Version 2.5.1 & Below. Product Usage: The CELLTRACKS AUTOPREP System is a general purpose laboratory instrument used with immunomagnetic reagents that capture and enrich target cells, and labeling reagents that differentiate cells in whole blood. Cell analyzers such as the CellTracks Analyzer II, CellSpotter System, flow cytometers or microscopes may be used for cell identification and enumeration. The system is fit for in vitro diagnostic use. In the CellT
Veridex, LLC has received a report that a patient's demographics were incorrectly associated with another patient's identification (PID) number. The purpose of this notification …
Veridex, LLC recalled CELLTRACKS ANALYZER II (Software Version 2.5.1 & Below. Product Usage: The CELLTRAC… - a moderate-severity action.
CELLTRACKS ANALYZER II (Software Version 2.5.1 & Below. Product Usage: The CELLTRAC… was recalled by Veridex, LLC in August 21, 2013. Reason: Veridex, LLC has received a report that a patient's demographics were incorrectly associated with another pat…. Check the official notice for the remedy. Verify recall #Z-1950-2013 with the FDA Devices before acting.
The recall
Veridex, LLC issued this moderate-severity FDA Devices recall-Veridex, LLC has received a report that a patient's demographics were incorrectly associated with another pat….
- Moderate
- severity level
- Class II
- classification
- August 21, 2013
- reported
Sourced from official FDA Devices enforcement records. Verify recall #Z-1950-2013 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
Recall Insight
This recall's affected scope, documented in the Recall Profile below, stays within a single-region or limited-distribution footprint. Distribution scope (see Distribution below) directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Medical Devices recalls over time
Where this recall sits in its category - 41,021 medical devices recalls on record
Of 88,244 recalls with an agency hazard classification, 19,068 are high severity, 64,652 moderate, and 4,524 low. A further 16,179 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Product description
CELLTRACKS ANALYZER II (Software Version 2.5.1 & Below. Product Usage: The CELLTRACKS AUTOPREP System is a general purpose laboratory instrument used with immunomagnetic reagents that capture and enrich target cells, and labeling reagents that differentiate cells in whole blood. Cell analyzers such as the CellTracks Analyzer II, CellSpotter System, flow cytometers or microscopes may be used for cell identification and enumeration. The system is fit for in vitro diagnostic use. In the CellTracks Analyzer II software 2.5.1 release notes Pub. No. J40136EN issued 2012-03-26 page 1, the document states that the edit patient sample information screen now includes a patient information section, a new feature unique to software 2.5.1. This section provides a drop down for patient race and cancer type. There are also separate fields for first and last name for both the physician and patient. There is no mention of the drop down patient ID box on the right side of the user interface which auto-populates previous sample data.
Reason for recall
Veridex, LLC has received a report that a patient's demographics were incorrectly associated with another patient's identification (PID) number. The purpose of this notification is to inform you of the issue and provide instructions to help prevent an occurrence on your CELLTRACKS AUTOPREP System and CELLTRACKS ANALYZER II.
Recall Profile
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-1950-2013 |
| Date reported | August 21, 2013 |
| Date initiated | December 28, 2012 |
| Recalling firm | Veridex, LLC |
| Firm location | Raritan, NJ |
| Affected scope | US: 70 Units; Ex-US 76 Units |
| Distribution | Worldwide Distribution - USA Nationwide and the countries of Australia, Canada, China, India, Japan, Singapore, England, France, Germany, Italy and Spain. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
- Check the recall number (Z-1950-2013) and product description against the item you own. Search the archive
- Confirm the current status with FDA Devices before acting, recall details can be updated.
- Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Where was the recalled product distributed? ▼
Should I stop using a recalled medication or medical device? ▼
Nationwide recalls with similar severity or category timing
Drug and device notices outside this firm and state, matched by severity timing and product category.
Similar severity · nearest report date
Same archive severity grade (Moderate), closest by report date among other firms and states.
Similar Medical Devices · nearest report date
Same product category, closest by report date among other firms and states (deduped against the severity set).
Compare this recall with GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VS… → · All Medical Devices recalls
Methodology · Corrections · Archive as of August 2026