FDA Devices recall · Reported May 6, 2026
Brand Name: HiResolution" Bionic Ear System Product Name: M Zn-Air Battery Pak Model/Catalog Number: CI-5501-110; CI-5501-120; CI-5501-130; CI-5501-140; CI-5501-150; CI-5501-190; CI-5501-240; CI-5501-250; CI-5501-260; CI-5501-270; CI-5501-280 Software Version: Not Applicable Product Description: The M Zn-Air Battery Pak is a disposable power source for the Naida" CI M and Sky CI" M Sound Processors. The type of disposable battery to be used with the M Zn-Air Battery Pak is a 675 power implan
The non-tamperproof battery door supplied with the sound processor fails to conform to IEC 60601-1-11:2015 by not requiring the use of a tool to replace the button/coin cell batte…
Advanced Bionics, LLC recalled Brand Name: HiResolution" Bionic Ear System Product Name: M Zn-Air Battery Pak Model/Ca… - a moderate-severity action.
Brand Name: HiResolution" Bionic Ear System Product Name: M Zn-Air Battery Pak Model/Ca… was recalled by Advanced Bionics, LLC in May 6, 2026. Reason: The non-tamperproof battery door supplied with the sound processor fails to conform to IEC 60601-1-11:2015 by…. Check the official notice for the remedy. Verify recall #Z-1950-2026 with the FDA Devices before acting.
The recall
Advanced Bionics, LLC issued this moderate-severity FDA Devices recall-The non-tamperproof battery door supplied with the sound processor fails to conform to IEC 60601-1-11:2015 by….
- Moderate
- severity level
- 45K units
- affected scope
- Class II
- classification
- May 6, 2026
- reported
Sourced from official FDA Devices enforcement records. Verify recall #Z-1950-2026 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
Recall Insight
This recall's affected scope, documented in the Recall Profile below, stays within a single-region or limited-distribution footprint. Distribution scope (see Distribution below) directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Medical Devices recalls over time
Where this recall sits in its category - 41,021 medical devices recalls on record
Of 88,244 recalls with an agency hazard classification, 19,068 are high severity, 64,652 moderate, and 4,524 low. A further 16,179 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Product description
Brand Name: HiResolution" Bionic Ear System Product Name: M Zn-Air Battery Pak Model/Catalog Number: CI-5501-110; CI-5501-120; CI-5501-130; CI-5501-140; CI-5501-150; CI-5501-190; CI-5501-240; CI-5501-250; CI-5501-260; CI-5501-270; CI-5501-280 Software Version: Not Applicable Product Description: The M Zn-Air Battery Pak is a disposable power source for the Naida" CI M and Sky CI" M Sound Processors. The type of disposable battery to be used with the M Zn-Air Battery Pak is a 675 power implant battery (P675). Regular 675 hearing instrument batteries will not be powerful enough for a cochlear implant. The suggested manufacturer and battery is PowerOne P675 Implant Plus battery. The M Zn-Air Battery Pak is available with non-tamperproof and tamperproof cartridges. Component: Not Applicable
Reason for recall
The non-tamperproof battery door supplied with the sound processor fails to conform to IEC 60601-1-11:2015 by not requiring the use of a tool to replace the button/coin cell battery.
Recall Profile
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Ongoing |
| Recall number | Z-1950-2026 |
| Date reported | May 6, 2026 |
| Date initiated | March 23, 2026 |
| Recalling firm | Advanced Bionics, LLC |
| Firm location | Valencia, CA |
| Affected scope | 45173 |
| Distribution | Worldwide - U.S Nationwide distribution including in the state of AK, AL, AR, AZ, CA, CO, CT, CD, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, T… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
- Check the recall number (Z-1950-2026) and product description against the item you own. Search the archive
- Confirm the current status with FDA Devices before acting, recall details can be updated.
- Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Where was the recalled product distributed? ▼
Should I stop using a recalled medication or medical device? ▼
Nationwide recalls with similar severity or category timing
Drug and device notices outside this firm and state, matched by severity timing and product category.
Similar severity · nearest report date
Same archive severity grade (Moderate), closest by report date among other firms and states.
Similar Medical Devices · nearest report date
Same product category, closest by report date among other firms and states (deduped against the severity set).
Compare this recall with GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VS… → · All Medical Devices recalls
Methodology · Corrections · Archive as of August 2026