A-Series BiPAP A 40, A-Series BiPAP A30 Continuous Ventilator, Non-life Supporting

Reported: July 21, 2021 Initiated: June 14, 2021 #Z-1958-2021

Product Description

A-Series BiPAP A 40, A-Series BiPAP A30 Continuous Ventilator, Non-life Supporting

Reason for Recall

The polyester-based polyurethane (PE-PUR) foam may degrade into particles which may enter the devices air pathway and be ingested or inhaled by the user, and the PE-PUR foam may off-gas certain chemicals.

Details

Recalling Firm
Philips Respironics, Inc.
Units Affected
72665
Distribution
Worldwide distribution including US Nationwide, Argentina, Australia, Brazil, Canada, China, France, Germany, Italy, Japan, Korea, and Latin America.
Location
Murrysville, PA

Frequently Asked Questions

What product was recalled?
A-Series BiPAP A 40, A-Series BiPAP A30 Continuous Ventilator, Non-life Supporting. Recalled by Philips Respironics, Inc.. Units affected: 72665.
Why was this product recalled?
The polyester-based polyurethane (PE-PUR) foam may degrade into particles which may enter the devices air pathway and be ingested or inhaled by the user, and the PE-PUR foam may off-gas certain chemicals.
Which agency issued this recall?
This recall was issued by the FDA Devices on July 21, 2021. Severity: Critical. Recall number: Z-1958-2021.