Severity
Moderate
FDA Devices recall · Reported July 18, 2012
The gravity steam, Ethylene Oxide (EO) and STERRAD 100S sterilization methods are being removed from the "Instructions for Use"; due to the firm's evaluation that identified origi…
Stryker Endoscopy recalled Stryker brand Arthroscope and Hardware Set Tray (sterilization tray); Model/Part Number: … - a moderate-severity action.
Stryker brand Arthroscope and Hardware Set Tray (sterilization tray); Model/Part Number: … was recalled by Stryker Endoscopy in July 18, 2012. Reason: The gravity steam, Ethylene Oxide (EO) and STERRAD 100S sterilization methods are being removed from the "Ins…. Check the official notice for the remedy. Verify recall #Z-1961-2012 with the FDA Devices before acting.
The recall
Stryker Endoscopy issued this moderate-severity FDA Devices recall-The gravity steam, Ethylene Oxide (EO) and STERRAD 100S sterilization methods are being removed from the "Ins….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1961-2012 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1961-2012) was formally reported on July 18, 2012, with the manufacturer initiating the action on June 19, 2012. It is classified under Moderate severity (Class II), with a current status of Terminated. Stryker Endoscopy is listed as the recalling firm, operating out of San Jose, CA. Federal records list the affected scope as 6388 units.
The documented reason for this recall is: The gravity steam, Ethylene Oxide (EO) and STERRAD 100S sterilization methods are being removed from the "Instructions for Use"; due to the firm's evaluation that identified original validation was not done at worst cas… Distribution data in the federal record shows the product reached: Worldwide Distribution--USA (nationwide). Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
6388 units
Related Recalls
6
3 from same agency
Stryker brand Arthroscope and Hardware Set Tray (sterilization tray); Model/Part Number: 233-032-116; Product is manufactured and distributed by Stryker Endoscopy, San Jose, CA The Arthroscope and Hardware Set Tray is used to hold and protect surgical devices during the sterilization process. Specifically this tray is used for arthroscopes and arthroscopic hardware devices. The tray consists of an interlocking tray and lid which are perforated to allow for the passage of sterilizing agent from outside the tray to the inside of the tray where the devices are held.
The gravity steam, Ethylene Oxide (EO) and STERRAD 100S sterilization methods are being removed from the "Instructions for Use"; due to the firm's evaluation that identified original validation was not done at worst case conditions.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-1961-2012 |
| Date reported | July 18, 2012 |
| Date initiated | June 19, 2012 |
| Recalling firm | Stryker Endoscopy |
| Firm location | San Jose, CA |
| Affected scope | 6388 units |
| Distribution | Worldwide Distribution--USA (nationwide) |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported July 18, 2012. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.