Terumo¿ Advanced Perfusion System 1: 220/240V AC, Advanced Perfusion System 1 The Terumo¿ Advanced Perfusion System 1 is indicated for use in extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures only when used by a qualified perfusionist who is experienced in the operation of Sarns " or similar equipment.

Reported: August 21, 2013 Initiated: July 26, 2013 #Z-1969-2013

Product Description

Terumo¿ Advanced Perfusion System 1: 220/240V AC, Advanced Perfusion System 1 The Terumo¿ Advanced Perfusion System 1 is indicated for use in extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures only when used by a qualified perfusionist who is experienced in the operation of Sarns " or similar equipment.

Reason for Recall

The operators manual does not fully describe the sequence of expected events that will occur when a specific operating condition exists. Specifically, if the user is adjusting the System 1 centrifugal pump speed via the Central Control Monitor at the same time an alarm condition occurs with configuration for a pump coast response, the labeling in the operators manual does not fully describe how the pump will react.

Details

Units Affected
1657
Distribution
Worldwide Distribution - USA (nationwide) including the countries of Australia, Dubai, Hong Kong, Indonesia, Singapore, China, Mexico, British Columbia, Taiwan, Thailand, Chile, Malaysia, Japan, Belgium, United Arab Emirates (UAE), South Korea, Manila and Canada.
Location
Ann Arbor, MI

Frequently Asked Questions

What product was recalled?
Terumo¿ Advanced Perfusion System 1: 220/240V AC, Advanced Perfusion System 1 The Terumo¿ Advanced Perfusion System 1 is indicated for use in extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures only when used by a qualified perfusionist who is experienced in the operation of Sarns " or similar equipment.. Recalled by Terumo Cardiovascular Systems Corporation. Units affected: 1657.
Why was this product recalled?
The operators manual does not fully describe the sequence of expected events that will occur when a specific operating condition exists. Specifically, if the user is adjusting the System 1 centrifugal pump speed via the Central Control Monitor at the same time an alarm condition occurs with configuration for a pump coast response, the labeling in the operators manual does not fully describe how the pump will react.
Which agency issued this recall?
This recall was issued by the FDA Devices on August 21, 2013. Severity: Moderate. Recall number: Z-1969-2013.