Severity
Moderate
FDA Devices recall · Reported July 15, 2015
On the Primus that includes the optional Chop / Lift Bar, there is a remote possibility that the Chop / Lift Bar could get accidently detached from the snap hook in specific use s…
Bte Technologies, Inc. recalled Primus (THE BTE WS30), model PrimusRS (PRRS) and Primus+ (PR30) a measuring exercise equi… - a moderate-severity action.
Primus (THE BTE WS30), model PrimusRS (PRRS) and Primus+ (PR30) a measuring exercise equi… was recalled by Bte Technologies, Inc. in July 15, 2015. Reason: On the Primus that includes the optional Chop / Lift Bar, there is a remote possibility that the Chop / Lift …. Check the official notice for the remedy. Verify recall #Z-1986-2015 with the FDA Devices before acting.
The recall
Bte Technologies, Inc. issued this moderate-severity FDA Devices recall-On the Primus that includes the optional Chop / Lift Bar, there is a remote possibility that the Chop / Lift ….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1986-2015 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1986-2015) was formally reported on July 15, 2015, with the manufacturer initiating the action on April 9, 2015. It is classified under Moderate severity (Class II), with a current status of Terminated. Bte Technologies, Inc. is listed as the recalling firm, operating out of Hanover, MD. Federal records list the affected scope as 510 Devices.
The documented reason for this recall is: On the Primus that includes the optional Chop / Lift Bar, there is a remote possibility that the Chop / Lift Bar could get accidently detached from the snap hook in specific use scenarios when the bar is rotated and twi… Distribution data in the federal record shows the product reached: Worldwide Distribution - Nationwide Distribution and to the countries of : Austria, Belgium, Canada, France, Germany, Greece, Hong Kong, Italy, India, Japan, South Korea, Mexico, P.R. China, Poland, Republic of South Af…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
510 Devices
Related Recalls
6
3 from same agency
Primus (THE BTE WS30), model PrimusRS (PRRS) and Primus+ (PR30) a measuring exercise equipment device designed to provide active and passive torque resistance to patient motion during rehabilitation therapy. The Primus that includes the optional Chop / Lift Bar.
On the Primus that includes the optional Chop / Lift Bar, there is a remote possibility that the Chop / Lift Bar could get accidently detached from the snap hook in specific use scenarios when the bar is rotated and twisted around the snap hook. The snap hook could open, resulting in possible injury to the user due to a fall.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-1986-2015 |
| Date reported | July 15, 2015 |
| Date initiated | April 9, 2015 |
| Recalling firm | Bte Technologies, Inc. |
| Firm location | Hanover, MD |
| Affected scope | 510 Devices |
| Distribution | Worldwide Distribution - Nationwide Distribution and to the countries of : Austria, Belgium, Canada, France, Germany, Greece, Hong Kong, Italy, India, Japan, South Korea, Mexico, P.R. China, Poland, Republic of South Africa, Romania, Saudi… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported July 15, 2015. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.