Patient Side Cart Battery Box used in conjunction with the da Vinci A surgical system. Intuitive Surgical Inc. Sunnyvale, CA 94086
Reported: July 16, 2014 Initiated: April 27, 2012 #Z-1994-2014
Product Description
Patient Side Cart Battery Box used in conjunction with the da Vinci A surgical system. Intuitive Surgical Inc. Sunnyvale, CA 94086
Reason for Recall
Patient Side Cart battery boxes may overheat in rare charging conditions, causing battery bulging.
Details
- Recalling Firm
- Intuitive Surgical, Inc.
- Units Affected
- 2606
- Distribution
- Worldwide Distribution.
- Location
- Sunnyvale, CA
Frequently Asked Questions
What product was recalled? ▼
Patient Side Cart Battery Box used in conjunction with the da Vinci A surgical system. Intuitive Surgical Inc. Sunnyvale, CA 94086. Recalled by Intuitive Surgical, Inc.. Units affected: 2606.
Why was this product recalled? ▼
Patient Side Cart battery boxes may overheat in rare charging conditions, causing battery bulging.
Which agency issued this recall? ▼
This recall was issued by the FDA Devices on July 16, 2014. Severity: Moderate. Recall number: Z-1994-2014.
Related Recalls
FDA Devices Moderate
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
FDA Devices Moderate
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
FDA Devices Moderate
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11