PushTracker E2 component utilized with the SmartDrive MX2+ Power Assist Device (Part Numbers: MX2-32K, MX2-32P, MX2-32PK, MX2-33D, MX2-33P, MX2-33PK, MX2-33S, MX2-150, MX2-167, MX2-32D); a wearable watch that communicates with a SmartDrive Power Assist Device via the SmartDrive MX2+ Application

Reported: July 5, 2023 Initiated: May 10, 2023 #Z-2004-2023

Product Description

PushTracker E2 component utilized with the SmartDrive MX2+ Power Assist Device (Part Numbers: MX2-32K, MX2-32P, MX2-32PK, MX2-33D, MX2-33P, MX2-33PK, MX2-33S, MX2-150, MX2-167, MX2-32D); a wearable watch that communicates with a SmartDrive Power Assist Device via the SmartDrive MX2+ Application

Reason for Recall

When multiple processes are running on the watch's Central Processing Unit (CPU), the application may fail unexpectedly. If this happens, the motor on the power assist device continues to run and the user may not be able to stop the device using tap gestures.

Details

Recalling Firm
Permobil
Units Affected
18,905 units
Distribution
Worldwide distribution.
Location
Lebanon, TN

Frequently Asked Questions

What product was recalled?
PushTracker E2 component utilized with the SmartDrive MX2+ Power Assist Device (Part Numbers: MX2-32K, MX2-32P, MX2-32PK, MX2-33D, MX2-33P, MX2-33PK, MX2-33S, MX2-150, MX2-167, MX2-32D); a wearable watch that communicates with a SmartDrive Power Assist Device via the SmartDrive MX2+ Application. Recalled by Permobil. Units affected: 18,905 units.
Why was this product recalled?
When multiple processes are running on the watch's Central Processing Unit (CPU), the application may fail unexpectedly. If this happens, the motor on the power assist device continues to run and the user may not be able to stop the device using tap gestures.
Which agency issued this recall?
This recall was issued by the FDA Devices on July 5, 2023. Severity: Moderate. Recall number: Z-2004-2023.