FDA Devices recall · Reported July 2, 2025
REMISOL Advance UDI-DI code: 13700962601874 All Software Versions: B92487; B92488; C24317; C28652; C37500; D04164; C44703; C57017; C69412; C69413; C73942; C73941; C88470; C88471 The Normand Info REMISOL Advance Data Manager collects and manages data and workflow for 128 connections (Beckman Coulter Instruments, automations, LIS&). It can also be connected to a host system. It provides data management capabilities such as automatic results validation, delta checking, reflex testing, quality
Due to a software issue, the software may incorrectly process the date of birth (DoB) received from the Laboratory Information System (LIS) and when the DoB is set to 01/01/1900 o…
Beckman Coulter Inc. recalled REMISOL Advance UDI-DI code: 13700962601874 All Software Versions: B92487; B92488; C24… - a moderate-severity action.
REMISOL Advance UDI-DI code: 13700962601874 All Software Versions: B92487; B92488; C24… was recalled by Beckman Coulter Inc. in July 2, 2025. Reason: Due to a software issue, the software may incorrectly process the date of birth (DoB) received from the Labor…. Check the official notice for the remedy. Verify recall #Z-2028-2025 with the FDA Devices before acting.
The recall
Beckman Coulter Inc. issued this moderate-severity FDA Devices recall-Due to a software issue, the software may incorrectly process the date of birth (DoB) received from the Labor….
- Moderate
- severity level
- Class II
- classification
- July 2, 2025
- reported
Sourced from official FDA Devices enforcement records. Verify recall #Z-2028-2025 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
Recall Insight
This recall's affected scope, documented in the Recall Profile below, stays within a single-region or limited-distribution footprint. Distribution scope (see Distribution below) directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Medical Devices recalls over time
Where this recall sits in its category - 41,021 medical devices recalls on record
Of 88,244 recalls with an agency hazard classification, 19,068 are high severity, 64,652 moderate, and 4,524 low. A further 16,179 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Product description
REMISOL Advance UDI-DI code: 13700962601874 All Software Versions: B92487; B92488; C24317; C28652; C37500; D04164; C44703; C57017; C69412; C69413; C73942; C73941; C88470; C88471 The Normand Info REMISOL Advance Data Manager collects and manages data and workflow for 128 connections (Beckman Coulter Instruments, automations, LIS&). It can also be connected to a host system. It provides data management capabilities such as automatic results validation, delta checking, reflex testing, quality control, results editing, archiving and restoration of patient results.
Reason for recall
Due to a software issue, the software may incorrectly process the date of birth (DoB) received from the Laboratory Information System (LIS) and when the DoB is set to 01/01/1900 or left empty, REMISOL Advance may incorrectly identify a patient as a newborn, affecting the execution of validation rules based on incorrect reference ranges. The test results are not changed by this software defect and are accurately displayed by REMISOL Advance; however, the evaluation flagging of the results may be affected. This issue may lead to wrong reference range calculations and can result in erroneous flagged results being auto validated and uploaded to LIS.
Recall Profile
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Ongoing |
| Recall number | Z-2028-2025 |
| Date reported | July 2, 2025 |
| Date initiated | April 8, 2025 |
| Recalling firm | Beckman Coulter Inc. |
| Firm location | Brea, CA |
| Affected scope | 1,805 systems |
| Distribution | Worldwide - U.S. Nationwide distribution including in the states of AK, AL,AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, … |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
- Check the recall number (Z-2028-2025) and product description against the item you own. Search the archive
- Confirm the current status with FDA Devices before acting, recall details can be updated.
- Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Where was the recalled product distributed? ▼
Should I stop using a recalled medication or medical device? ▼
Nationwide recalls with similar severity or category timing
Drug and device notices outside this firm and state, matched by severity timing and product category.
Similar severity · nearest report date
Same archive severity grade (Moderate), closest by report date among other firms and states.
Similar Medical Devices · nearest report date
Same product category, closest by report date among other firms and states (deduped against the severity set).
Compare this recall with GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VS… → · All Medical Devices recalls
Methodology · Corrections · Archive as of August 2026