Biograph mCT-S(64) 3R, MATERIAL NUMBER 10248669 The Siemens Biograph TruePoint systems are combined X-Ray Computed Tomography (CT) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information.
Reported: June 29, 2016 Initiated: April 22, 2016 #Z-2032-2016
Product Description
Biograph mCT-S(64) 3R, MATERIAL NUMBER 10248669 The Siemens Biograph TruePoint systems are combined X-Ray Computed Tomography (CT) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information.
Reason for Recall
Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration.
Details
- Recalling Firm
- Siemens Medical Solutions USA Inc.
- Units Affected
- 51 units
- Distribution
- Worldwide Distribution - US Nationwide
- Location
- Knoxville, TN
Frequently Asked Questions
What product was recalled? ▼
Biograph mCT-S(64) 3R, MATERIAL NUMBER 10248669 The Siemens Biograph TruePoint systems are combined X-Ray Computed Tomography (CT) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information.. Recalled by Siemens Medical Solutions USA Inc.. Units affected: 51 units.
Why was this product recalled? ▼
Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration.
Which agency issued this recall? ▼
This recall was issued by the FDA Devices on June 29, 2016. Severity: Moderate. Recall number: Z-2032-2016.
Related Recalls
FDA Devices Moderate
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
FDA Devices Moderate
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
FDA Devices Moderate
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11