Radiation therapy software, RayStation standalone software treatment planning system. Model numbers: 8B, 8B SP1, 8B SP2, 9A, 9B, 9B SP1, 10A, 10A SP1 and 10B. Software versions: 8.1, 9.0, 9.1, 9.2, 10.0, 10.1.

Reported: July 5, 2023 Initiated: January 22, 2021 #Z-2039-2023

Product Description

Radiation therapy software, RayStation standalone software treatment planning system. Model numbers: 8B, 8B SP1, 8B SP2, 9A, 9B, 9B SP1, 10A, 10A SP1 and 10B. Software versions: 8.1, 9.0, 9.1, 9.2, 10.0, 10.1.

Reason for Recall

A software bug affecting results when using deep learning (DL) segmentation followed by geometrical functions: Create wall function, Expand Contract function or ROI algebra function, the results are incorrect. If dose statistics for an incorrectly created ROI are used to determine if a plan is acceptable for treatment, this could potentially lead to an inappropriate plan being approved. The error could lead to a more conservative or more liberal plan than intended.

Details

Recalling Firm
RaySearch America Inc
Units Affected
5 systems
Distribution
US: AZ, CA, GA & WA
Location
New York, NY

Frequently Asked Questions

What product was recalled?
Radiation therapy software, RayStation standalone software treatment planning system. Model numbers: 8B, 8B SP1, 8B SP2, 9A, 9B, 9B SP1, 10A, 10A SP1 and 10B. Software versions: 8.1, 9.0, 9.1, 9.2, 10.0, 10.1.. Recalled by RaySearch America Inc. Units affected: 5 systems.
Why was this product recalled?
A software bug affecting results when using deep learning (DL) segmentation followed by geometrical functions: Create wall function, Expand Contract function or ROI algebra function, the results are incorrect. If dose statistics for an incorrectly created ROI are used to determine if a plan is acceptable for treatment, this could potentially lead to an inappropriate plan being approved. The error could lead to a more conservative or more liberal plan than intended.
Which agency issued this recall?
This recall was issued by the FDA Devices on July 5, 2023. Severity: Moderate. Recall number: Z-2039-2023.