Severity
Moderate
FDA Devices recall · Reported July 9, 2025
a software anomaly allows test results to be reported using Micro Tip and Micro Slide diluent packs that the system treats as expired even when the Use Expired Reagents setting…
Ortho-Clinical Diagnostics, Inc. recalled Brand Name: VITROS; Product Name: VITROS XT 7600 Integrated System; Model/Catalog Numbe… - a moderate-severity action.
Brand Name: VITROS; Product Name: VITROS XT 7600 Integrated System; Model/Catalog Numbe… was recalled by Ortho-Clinical Diagnostics, Inc. in July 9, 2025. Reason: a software anomaly allows test results to be reported using Micro Tip and Micro Slide diluent packs that the…. Check the official notice for the remedy. Verify recall #Z-2039-2025 with the FDA Devices before acting.
The recall
Ortho-Clinical Diagnostics, Inc. issued this moderate-severity FDA Devices recall-a software anomaly allows test results to be reported using Micro Tip and Micro Slide diluent packs that the….
Sourced from official FDA Devices enforcement records. Verify recall #Z-2039-2025 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-2039-2025) was formally reported on July 9, 2025, with the manufacturer initiating the action on May 9, 2025. It is classified under Moderate severity (Class II), with a current status of Ongoing. Ortho-Clinical Diagnostics, Inc. is listed as the recalling firm, operating out of Rochester, NY. Federal records list the affected scope as 1958 units (1019 US, 939 OUS).
The documented reason for this recall is: a software anomaly allows test results to be reported using Micro Tip and Micro Slide diluent packs that the system treats as expired even when the Use Expired Reagents setting is not enabled if a shelf expiration d… Distribution data in the federal record shows the product reached: Worldwide Distribution: US (Nationwide) and OUS (International) to countries of: Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, …. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
1958 units (1019 US, 939 OUS)
Related Recalls
6
3 from same agency
Brand Name: VITROS; Product Name: VITROS XT 7600 Integrated System; Model/Catalog Number: (1) 6844461, (2) 6272222; Software Version: version 3.2 to version 3.8.3 (all available software versions); Product Description: In vitro diagnostic chemistry analyzer.
a software anomaly allows test results to be reported using Micro Tip and Micro Slide diluent packs that the system treats as expired even when the Use Expired Reagents setting is not enabled if a shelf expiration date is omitted during manual loading. The issue was identified by a customer complaint, in which a CREA Urine test was processed with an RE (Reagent Expired) code, even though the diluent pack was not expired and the use-expired-reagents setting was disabled. This can result in erroneous test results and/or delaying test results.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Ongoing |
| Recall number | Z-2039-2025 |
| Date reported | July 9, 2025 |
| Date initiated | May 9, 2025 |
| Recalling firm | Ortho-Clinical Diagnostics, Inc. |
| Firm location | Rochester, NY |
| Affected scope | 1958 units (1019 US, 939 OUS) |
| Distribution | Worldwide Distribution: US (Nationwide) and OUS (International) to countries of: Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, Norway, Portugal, Ru… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported July 9, 2025. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.