FDA Devices recall · Reported May 13, 2026
LUX-DX II, Arrythmia detector and alarm, Model M302, with SERVER SW LATITUDE DRAGON US, Model 6460, Version 7.5
For some patients upgraded from a LUX-Dx M301 device to a new Model M302 or M312, the new ICM device is not collecting PVC Burden data (both M302/M312) or monitoring for Bradyarrh…
Boston Scientific Corporation recalled LUX-DX II, Arrythmia detector and alarm, Model M302, with SERVER SW LATITUDE DRAGON US, M… - a moderate-severity action.
LUX-DX II, Arrythmia detector and alarm, Model M302, with SERVER SW LATITUDE DRAGON US, M… was recalled by Boston Scientific Corporation in May 13, 2026. Reason: For some patients upgraded from a LUX-Dx M301 device to a new Model M302 or M312, the new ICM device is not c…. Check the official notice for the remedy. Verify recall #Z-2048-2026 with the FDA Devices before acting.
The recall
Boston Scientific Corporation issued this moderate-severity FDA Devices recall-For some patients upgraded from a LUX-Dx M301 device to a new Model M302 or M312, the new ICM device is not c….
- Moderate
- severity level
- 2 units
- affected scope
- Class II
- classification
- May 13, 2026
- reported
Sourced from official FDA Devices enforcement records. Verify recall #Z-2048-2026 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
Recall Insight
This recall's affected scope, documented in the Recall Profile below, stays within a single-region or limited-distribution footprint. Distribution scope (see Distribution below) directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Medical Devices recalls over time
Where this recall sits in its category - 41,021 medical devices recalls on record
Of 88,244 recalls with an agency hazard classification, 19,068 are high severity, 64,652 moderate, and 4,524 low. A further 16,179 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Product description
LUX-DX II, Arrythmia detector and alarm, Model M302, with SERVER SW LATITUDE DRAGON US, Model 6460, Version 7.5
Reason for recall
For some patients upgraded from a LUX-Dx M301 device to a new Model M302 or M312, the new ICM device is not collecting PVC Burden data (both M302/M312) or monitoring for Bradyarrhythmia and Pause (M312 only). In addition, for these devices that are not being monitored, the LATITUDE Clarity programming screens have discrepant information, indicating that the monitoring is enabled for these features in one location and not enabled in another location.
Recall Profile
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Ongoing |
| Recall number | Z-2048-2026 |
| Date reported | May 13, 2026 |
| Date initiated | March 30, 2026 |
| Recalling firm | Boston Scientific Corporation |
| Firm location | Saint Paul, MN |
| Affected scope | 2 units |
| Distribution | Worldwide distribution - US Nationwide and the country of Belgium. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
- Check the recall number (Z-2048-2026) and product description against the item you own. Search the archive
- Confirm the current status with FDA Devices before acting, recall details can be updated.
- Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Where was the recalled product distributed? ▼
Should I stop using a recalled medication or medical device? ▼
Nationwide recalls with similar severity or category timing
Drug and device notices outside this firm and state, matched by severity timing and product category.
Similar severity · nearest report date
Same archive severity grade (Moderate), closest by report date among other firms and states.
Similar Medical Devices · nearest report date
Same product category, closest by report date among other firms and states (deduped against the severity set).
Compare this recall with GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VS… → · All Medical Devices recalls
Methodology · Corrections · Archive as of August 2026