SHAPE ARM, SINGLE, an articulating, arm bracket that is attached to the service head of boom systems which are used for supporting and positioning equipment in operating rooms and patient care areas. The product comes as a component of kit part number 0682400253 and the Arm part number in the kit is P17085. . The kit, when assembled, supports a flat-panel video monitor that is attached to one of two types of Stryker Communications boom systems, EDS or FLEXiS.

Reported: July 23, 2014 Initiated: March 17, 2014 #Z-2052-2014

Product Description

SHAPE ARM, SINGLE, an articulating, arm bracket that is attached to the service head of boom systems which are used for supporting and positioning equipment in operating rooms and patient care areas. The product comes as a component of kit part number 0682400253 and the Arm part number in the kit is P17085. . The kit, when assembled, supports a flat-panel video monitor that is attached to one of two types of Stryker Communications boom systems, EDS or FLEXiS.

Reason for Recall

During the use of the SHAPE Arm there is the potential for the monitor mount to come off of the SHAPE Arm if the lock washer and screw are not present.

Details

Recalling Firm
Stryker Communications
Units Affected
953 STRYKER SHAPE ARMS of all types
Distribution
Worldwide Distribution. US nationwide including Puerto Rico, China, Philippines, Mexico, Korea, Japan, Colombia, Chile, Canada, Brazil, and Australia
Location
Flower Mound, TX

Frequently Asked Questions

What product was recalled?
SHAPE ARM, SINGLE, an articulating, arm bracket that is attached to the service head of boom systems which are used for supporting and positioning equipment in operating rooms and patient care areas. The product comes as a component of kit part number 0682400253 and the Arm part number in the kit is P17085. . The kit, when assembled, supports a flat-panel video monitor that is attached to one of two types of Stryker Communications boom systems, EDS or FLEXiS.. Recalled by Stryker Communications. Units affected: 953 STRYKER SHAPE ARMS of all types.
Why was this product recalled?
During the use of the SHAPE Arm there is the potential for the monitor mount to come off of the SHAPE Arm if the lock washer and screw are not present.
Which agency issued this recall?
This recall was issued by the FDA Devices on July 23, 2014. Severity: Moderate. Recall number: Z-2052-2014.