FDA Devices recall · Reported August 22, 2012
Stryker Neptune Bronze Rover, Part 0700-007-000 Neptune Bronze Rover Waste Management System Instructions For Use. Intended for use in the surgical environment or physician's office to collect and dispose of surgical fluid waste.
Stryker has received two reports of serious injury as a result of tissue damage associated with the use of the Neptune 2, including an event in which one customer connected the Ne…
Stryker Instruments Div. of Stryker Corporation recalled Stryker Neptune Bronze Rover, Part 0700-007-000 Neptune Bronze Rover Waste Management Sy… - a critical-severity action.
Stryker Neptune Bronze Rover, Part 0700-007-000 Neptune Bronze Rover Waste Management Sy… was recalled by Stryker Instruments Div. of Stryker Corporation in August 22, 2012. Reason: Stryker has received two reports of serious injury as a result of tissue damage associated with the use of th…. Check the official notice for the remedy. Verify recall #Z-2063-2012 with the FDA Devices before acting.
The recall
Stryker Instruments Div. of Stryker Corporation issued this critical-severity FDA Devices recall-Stryker has received two reports of serious injury as a result of tissue damage associated with the use of th….
- Critical
- severity level
- 1K units
- affected scope
- Class I
- classification
- August 22, 2012
- reported
Sourced from official FDA Devices enforcement records. Verify recall #Z-2063-2012 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
Recall Insight
This recall's affected scope, documented in the Recall Profile below, stays within a single-region or limited-distribution footprint. Distribution scope (see Distribution below) directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Medical Devices recalls over time
Where this recall sits in its category - 41,021 medical devices recalls on record
Of 88,244 recalls with an agency hazard classification, 19,068 are high severity, 64,652 moderate, and 4,524 low. A further 16,179 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified high severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Product description
Stryker Neptune Bronze Rover, Part 0700-007-000 Neptune Bronze Rover Waste Management System Instructions For Use. Intended for use in the surgical environment or physician's office to collect and dispose of surgical fluid waste.
Reason for recall
Stryker has received two reports of serious injury as a result of tissue damage associated with the use of the Neptune 2, including an event in which one customer connected the Neptune 2 System to a passive chest drainage tube post operatively, resulting in a fatality. The current IFU did not specifically warn against connecting all Neptune Waste Management devices, which is a high vacuum/high flow device, to a passive drainage tube. June 12, 2012, customers were instructed to review the revised IFU, distribute to affected departments, and educate users of the Neptune on this warning. Warning Labels, which are pending distribution, are required on all Neptune devices. Further investigation found the Neptune 1 Silver, Neptune 2 Ultra (120V) and Neptune 2 Ultra (230V) lacked FDA 510(k) clearance and therefore are not legally marketed without determination of their safety and effectiveness. FDA advises Neptune 1 Silver, Neptune 2 Ultra (120V) and Neptune 2 Ultra (230V) not be used. However, customers who do not have an alternative device to use should weigh the risks and benefits associated with continued use of these devices. If customers choose to continue use of the Neptune Silver, Neptune 2 Ultra (120V) or Neptune 2 (230V), they must complete a Certificate of Medical Necessity and return it to Stryker by October 12, 2012.
Recall Profile
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Critical (Class I) |
| Status | Terminated |
| Recall number | Z-2063-2012 |
| Date reported | August 22, 2012 |
| Date initiated | June 8, 2012 |
| Recalling firm | Stryker Instruments Div. of Stryker Corporation |
| Firm location | Portage, MI |
| Affected scope | 1075 |
| Distribution | Worldwide Distribution -- USA (nationwide) and country of: Canada. . |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
- Check the recall number (Z-2063-2012) and product description against the item you own. Search the archive
- Confirm the current status with FDA Devices before acting, recall details can be updated.
- This is a Class I (high-risk) recall, stop using the product immediately and follow the disposal or return instructions. What to do next
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Where was the recalled product distributed? ▼
Should I stop using a recalled medication or medical device? ▼
Nationwide recalls with similar severity or category timing
Drug and device notices outside this firm and state, matched by severity timing and product category.
Similar severity · nearest report date
Same archive severity grade (Critical), closest by report date among other firms and states.
Similar Medical Devices · nearest report date
Same product category, closest by report date among other firms and states (deduped against the severity set).
Compare this recall with GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VS… → · All Medical Devices recalls
Methodology · Corrections · Archive as of August 2026