Severity
Moderate
FDA Devices recall · Reported August 1, 2012
software automatically switches the lot when a vancomycin (VANC) or valproic acid (VALP) reagent is depleted or the operator manually changes the lot in use causing lack of verifi…
Ortho-Clinical Diagnostics recalled VITROS 4600 Chemistry System Software For use in the in-vitro quantitative, semi quan… - a moderate-severity action.
VITROS 4600 Chemistry System Software For use in the in-vitro quantitative, semi quan… was recalled by Ortho-Clinical Diagnostics in August 1, 2012. Reason: software automatically switches the lot when a vancomycin (VANC) or valproic acid (VALP) reagent is depleted …. Check the official notice for the remedy. Verify recall #Z-2067-2012 with the FDA Devices before acting.
The recall
Ortho-Clinical Diagnostics issued this moderate-severity FDA Devices recall-software automatically switches the lot when a vancomycin (VANC) or valproic acid (VALP) reagent is depleted ….
Sourced from official FDA Devices enforcement records. Verify recall #Z-2067-2012 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-2067-2012) was formally reported on August 1, 2012, with the manufacturer initiating the action on May 21, 2012. It is classified under Moderate severity (Class II), with a current status of Terminated. Ortho-Clinical Diagnostics is listed as the recalling firm, operating out of Rochester, NY. Federal records list the affected scope as 77.
The documented reason for this recall is: software automatically switches the lot when a vancomycin (VANC) or valproic acid (VALP) reagent is depleted or the operator manually changes the lot in use causing lack of verification of proper quality assessment. Distribution data in the federal record shows the product reached: Worldwide Distribution - USA (nationwide) and the countries of Australia, New Zealand, Brazil, Canada, Chile, Argentina, Paraquay, Ecuador, Colombia, India, Japan, Mexico, Panama, Singapore, Venezuela, UK, Northern Euro…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
77
Related Recalls
6
3 from same agency
VITROS 4600 Chemistry System Software For use in the in-vitro quantitative, semi quantitative, and qualitative measurement of variety of analytes of clinical interest.
software automatically switches the lot when a vancomycin (VANC) or valproic acid (VALP) reagent is depleted or the operator manually changes the lot in use causing lack of verification of proper quality assessment.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-2067-2012 |
| Date reported | August 1, 2012 |
| Date initiated | May 21, 2012 |
| Recalling firm | Ortho-Clinical Diagnostics |
| Firm location | Rochester, NY |
| Affected scope | 77 |
| Distribution | Worldwide Distribution - USA (nationwide) and the countries of Australia, New Zealand, Brazil, Canada, Chile, Argentina, Paraquay, Ecuador, Colombia, India, Japan, Mexico, Panama, Singapore, Venezuela, UK, Northern Europe, France, Germany,… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported August 1, 2012. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.