Centurion Convenience Kits, containing BD Alaris Pump Infusion Sets. BD code 11522558 (Centurion code 110931) contained inside kits DYNJ52509A and DYNJ52510A. Product Usage: The Centurion Primary Set Pack and Primary Warmer Pack Convenience Kit helps continuously or intermittently deliver fluids, medications, blood and blood products to adult, pediatric or neonatal patients.

Reported: August 21, 2019 Initiated: May 10, 2019 #Z-2067-2019 4890 units

Medline Industries Inc issued this FDA Devices recall on August 21, 2019. Classified as Critical severity (Class I). Approximately 4890 units are affected. The recall was issued because: This recall is a sub recall to BD s recent Class I recall for the Alaris Pump Infusion Set due to an issue with the con…. This recall notice is sourced from official FDA Devices enforcement records. Below you will find the complete product description, hazard information, remedy instructions, and related recalls from the same manufacturer or product category.

Recall Insight

This FDA Devices action (record #Z-2067-2019) was formally reported on August 21, 2019, with the manufacturer initiating the action on May 10, 2019. It is classified under Critical severity (Class I), with a current status of Terminated. Medline Industries Inc is listed as the recalling firm, operating out of Northfield, IL. Federal records indicate 4890 units are affected.

The documented reason for this recall is: This recall is a sub recall to BD s recent Class I recall for the Alaris Pump Infusion Set due to an issue with the consistency of the inner lumen thickness of the silicone tubing segment which directly impacts amount o… Distribution data in the federal record shows the product reached: One distributor located in Illinois.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

To put this record in context, PlainRecalls indexes 83,949 recalls across the FDA, CPSC, NHTSA and USDA FSIS going back to 1995. Within the same product category, the database holds 6 closely related recalls, of which 6 were also issued by FDA Devices. That clustering is a signal — repeated actions in a narrow category often indicate a systemic quality-control issue, a supplier-wide contamination, or a design defect that has propagated across product lines. This recall is roughly 7 years old; older recalls can remain relevant because many units enter resale, rental, and secondary-market channels where the original warning never reaches the end user. Always cross-check the recall number against the official agency page before relying on any summary.

Recall Distribution by Severity Class

Severity1Class I (Critical)Class II (Moderate)Class III (Low)
Recall Distribution by Severity Class

Severity

Critical

Units Affected

4890

Related Recalls

6

6 from same agency

Recall Progress (industry avg ~60%) 60.0%

Product Description

Centurion Convenience Kits, containing BD Alaris Pump Infusion Sets. BD code 11522558 (Centurion code 110931) contained inside kits DYNJ52509A and DYNJ52510A. Product Usage: The Centurion Primary Set Pack and Primary Warmer Pack Convenience Kit helps continuously or intermittently deliver fluids, medications, blood and blood products to adult, pediatric or neonatal patients.

Reason for Recall

This recall is a sub recall to BD s recent Class I recall for the Alaris Pump Infusion Set due to an issue with the consistency of the inner lumen thickness of the silicone tubing segment which directly impacts amount of occlusion force necessary to fully seal the tube. This defect prevents the pump from complete occlusion and results in the potential for over-infusion of drug product into the patient.

Details

Recalling Firm
Medline Industries Inc
Units Affected
4890
Distribution
One distributor located in Illinois.
Location
Northfield, IL

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Critical (Class I)
Status Terminated
Recall number Z-2067-2019
Date reported August 21, 2019
Date initiated May 10, 2019
Recalling firm Medline Industries Inc
Units affected 4890
Distribution One distributor located in Illinois.

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

Scale of Impact

4890 units affected — limited or regional distribution scale.

Regional (<10K units) ✓ This recall
Multi-state (10K – 100K units)
Large-scale (100K – 1M units)
Massive (≥1M units)

Bracket cutoffs follow federal recall-disclosure conventions; bar widths scale linearly within each bracket. Source: PlainRecalls analysis of U.S. Food and Drug Administration filings.

Frequently Asked Questions

What product was recalled?
Centurion Convenience Kits, containing BD Alaris Pump Infusion Sets. BD code 11522558 (Centurion code 110931) contained inside kits DYNJ52509A and DYNJ52510A. Product Usage: The Centurion Primary Set Pack and Primary Warmer Pack Convenience Kit helps continuously or intermittently deliver fluids, medications, blood and blood products to adult, pediatric or neonatal patients.. Recalled by Medline Industries Inc. Units affected: 4890.
Why was this product recalled?
This recall is a sub recall to BD s recent Class I recall for the Alaris Pump Infusion Set due to an issue with the consistency of the inner lumen thickness of the silicone tubing segment which directly impacts amount of occlusion force necessary to fully seal the tube. This defect prevents the pump from complete occlusion and results in the potential for over-infusion of drug product into the patient.
Which agency issued this recall?
This recall was issued by the FDA Devices on August 21, 2019. Severity: Critical. Recall number: Z-2067-2019.
Where was the recalled product distributed?
Distribution: One distributor located in Illinois..
How do I check if my product is affected by a recall?
Check the product description and recall number (Z-2067-2019) against your product. Visit the official FDA Devices website for the most current information. You can also use our Recall Checker tool to search by product name or brand.
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Recall Context

Product recalls are issued when a manufacturer, distributor, or federal agency determines that a product poses a safety risk to consumers. This recall is classified as critical severity, meaning the product carries a reasonable probability of serious adverse health consequences or death. Across PlainRecalls, we track 83,000+ recalls from FDA, CPSC, and NHTSA to help consumers stay informed and act quickly when safety issues arise.

Nearby Recalls in This Category

Other recalls in the same product category — useful for spotting patterns across the same defect class or manufacturer.

Compare this recall with Endo-Model Replacement Plateau; Item Number: 15-0027/11; →

Data Sources

Data as of 2025. Source: FDA, CPSC, NHTSA, USDA FSIS federal recall databases.

  • Source: FDA — Food and Drug Administration, openFDA Enforcement API (food, drug, and medical device recalls)
  • Source: CPSC — Consumer Product Safety Commission Recalls API (consumer product recalls and hazards)
  • Source: NHTSA — National Highway Traffic Safety Administration Recalls API (vehicle safety recalls)
  • Source: USDA FSIS — Food Safety and Inspection Service (meat, poultry, and egg product recalls)

Recall information is sourced from official federal agency databases. Always verify recall details with the issuing agency for the most current status. This information is for research and awareness purposes only.

All federal data sources used on this page

Source: Federal recall agencies (FDA, CPSC, NHTSA, USDA FSIS) Aggregated multi-agency recall feeds · 2024 Recall data normalized across federal agency feeds; severity classifications follow each agency's own taxonomy (FDA Class I/II/III; CPSC, NHTSA, USDA FSIS).