FDA Devices recall · Reported July 30, 2014
DATEX-OHMEDA S/5 SINGLE-WIDTH AIRWAY MODULE, E-MINIC AND ACCESSORIES. DATEX-OHMEDA S/5TM FM WITH L-FICU04 ANDL-FICUO4A SOFTWARE AND N-FCREC MODULE **(OUS only). E-miniC and accessories is indicated for monitoring C02 and respiration rate of all hospital patients.
Failure of the CO2 detector in Single-width Airway and Extension modules, which may cause a slow continuous decrease of measured Et-/FiCO2 values. Incorrect Et CO2/Fi CO2 value ma…
Ge Healthcare, LLC recalled DATEX-OHMEDA S/5 SINGLE-WIDTH AIRWAY MODULE, E-MINIC AND ACCESSORIES. DATEX-OHMEDA S/… - a critical-severity action.
DATEX-OHMEDA S/5 SINGLE-WIDTH AIRWAY MODULE, E-MINIC AND ACCESSORIES. DATEX-OHMEDA S/… was recalled by Ge Healthcare, LLC in July 30, 2014. Reason: Failure of the CO2 detector in Single-width Airway and Extension modules, which may cause a slow continuous d…. Check the official notice for the remedy. Verify recall #Z-2068-2014 with the FDA Devices before acting.
The recall
Ge Healthcare, LLC issued this critical-severity FDA Devices recall-Failure of the CO2 detector in Single-width Airway and Extension modules, which may cause a slow continuous d….
- Critical
- severity level
- Class I
- classification
- July 30, 2014
- reported
Sourced from official FDA Devices enforcement records. Verify recall #Z-2068-2014 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
Recall Insight
This recall's affected scope, documented in the Recall Profile below, stays within a single-region or limited-distribution footprint. Distribution scope (see Distribution below) directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Medical Devices recalls over time
Where this recall sits in its category - 41,021 medical devices recalls on record
Of 88,244 recalls with an agency hazard classification, 19,068 are high severity, 64,652 moderate, and 4,524 low. A further 16,179 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified high severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Product description
DATEX-OHMEDA S/5 SINGLE-WIDTH AIRWAY MODULE, E-MINIC AND ACCESSORIES. DATEX-OHMEDA S/5TM FM WITH L-FICU04 ANDL-FICUO4A SOFTWARE AND N-FCREC MODULE **(OUS only). E-miniC and accessories is indicated for monitoring C02 and respiration rate of all hospital patients.
Reason for recall
Failure of the CO2 detector in Single-width Airway and Extension modules, which may cause a slow continuous decrease of measured Et-/FiCO2 values. Incorrect Et CO2/Fi CO2 value may impair clinical decision making for both mechanically and spontaneously ventilated patients because of incorrect low CO2 values.
Recall Profile
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Critical (Class I) |
| Status | Terminated |
| Recall number | Z-2068-2014 |
| Date reported | July 30, 2014 |
| Date initiated | June 11, 2014 |
| Recalling firm | Ge Healthcare, LLC |
| Firm location | Waukesha, WI |
| Affected scope | 2,948 (761 units US, 2,187 units OUS) |
| Distribution | Worldwide Distribution - USA nationwide, ARGENTINA, AUSTRALIA, AUSTRIA, BAHRAIN, BRAZIL, CANADA, CHILE, CHINA, Croatia, CZECH REPUBLIC, DENMARK, EGYPT, EL SALVADOR, FINLAND, FRANCE, GERMANY, GREECE, GUATEMALA, HONG KONG, INDIA, IRAQ, ISRAE… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
- Check the recall number (Z-2068-2014) and product description against the item you own. Search the archive
- Confirm the current status with FDA Devices before acting, recall details can be updated.
- This is a Class I (high-risk) recall, stop using the product immediately and follow the disposal or return instructions. What to do next
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Where was the recalled product distributed? ▼
Should I stop using a recalled medication or medical device? ▼
Nationwide recalls with similar severity or category timing
Drug and device notices outside this firm and state, matched by severity timing and product category.
Similar severity · nearest report date
Same archive severity grade (Critical), closest by report date among other firms and states.
Similar Medical Devices · nearest report date
Same product category, closest by report date among other firms and states (deduped against the severity set).
Compare this recall with GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VS… → · All Medical Devices recalls
Methodology · Corrections · Archive as of August 2026