Severity
Moderate
FDA Devices recall · Reported May 24, 2017
Diamond particulates may break off the burr and fall into the surgical site. If the particles are not detected and are not removed from the surgical site a foreign body reaction …
Aesculap Implant Systems LLC recalled GP161R; ELAN 4 1-RING DIAMOND BURR COARSE D2.0; GP162R; ELAN 4 1-RING DIAMOND BURR+ CO… - a moderate-severity action.
GP161R; ELAN 4 1-RING DIAMOND BURR COARSE D2.0; GP162R; ELAN 4 1-RING DIAMOND BURR+ CO… was recalled by Aesculap Implant Systems LLC in May 24, 2017. Reason: Diamond particulates may break off the burr and fall into the surgical site. If the particles are not detect…. Check the official notice for the remedy. Verify recall #Z-2087-2017 with the FDA Devices before acting.
The recall
Aesculap Implant Systems LLC issued this moderate-severity FDA Devices recall-Diamond particulates may break off the burr and fall into the surgical site. If the particles are not detect….
Sourced from official FDA Devices enforcement records. Verify recall #Z-2087-2017 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-2087-2017) was formally reported on May 24, 2017, with the manufacturer initiating the action on March 28, 2017. It is classified under Moderate severity (Class II), with a current status of Terminated. Aesculap Implant Systems LLC is listed as the recalling firm, operating out of Center Valley, PA. Federal records list the affected scope as Total of all burrs - 608 units.
The documented reason for this recall is: Diamond particulates may break off the burr and fall into the surgical site. If the particles are not detected and are not removed from the surgical site a foreign body reaction may occur. These particulates could also… Distribution data in the federal record shows the product reached: US Nationwide distribution - CA, IL, NY, PA, TN, TX. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
Total of all burrs - 608 units
Related Recalls
6
3 from same agency
GP161R; ELAN 4 1-RING DIAMOND BURR COARSE D2.0; GP162R; ELAN 4 1-RING DIAMOND BURR+ COARSE D2.0; GP163R; ELAN 4 1-RING DIAMOND BURR COARSE D2,3; GP164R; ELAN 4 1-RING DIAMOND BURR+ COARSE D2,3; GP165R; ELAN 4 1-RING DIAMOND BURR COARSE D3,0; GP166R; ELAN 4 1-RING DIAMOND BURR+ COARSE D3,0; GP167R; ELAN 4 1-RING DIAMOND BURR++ COARSE D3,0; GP168R; ELAN 4 1-RING DIAMOND BURR COARSE D4,0; GP169R; ELAN 4 1-RING DIAMOND BURR COARSE D5,0; GP170R; ELAN 4 1-RING DIAMOND BURR COARSE D6,0; GP173R; ELAN 4 1-RING DIAMOND BURR X-COARSE D3,0; GP174R; ELAN 4 1-RING DIAMOND BURR X-COARSE D4,0; GP175R; ELAN 4 1-RING DIAMOND BURR X-COARSE D5,0; GP176R; ELAN 4 1-RING DIAMOND BURR X-COARSE D6,0; GP177R; ELAN 4 1-RING DIAMOND BURR X-COARSE D7,0; GP328R; ELAN 4 2-RING DIAMOND BURR X-COARSE D4.0; GP329R; ELAN 4 2-RING DIAMOND BURR X-COARSE D5.0; GP330R; ELAN 4 2-RING DIAMOND BURR X-COARSE D6.0; Product Usage: The ELAN 4 motor system and the Hi-Line XS handpiece are intended for high speed cutting, sawing, drilling and manipulation of soft tissue and bone in the fields of Spine, ENT, Neuro, and Maxillofacial Surgery.
Diamond particulates may break off the burr and fall into the surgical site. If the particles are not detected and are not removed from the surgical site a foreign body reaction may occur. These particulates could also lead to mechanically induced soft tissue damage.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-2087-2017 |
| Date reported | May 24, 2017 |
| Date initiated | March 28, 2017 |
| Recalling firm | Aesculap Implant Systems LLC |
| Firm location | Center Valley, PA |
| Affected scope | Total of all burrs - 608 units |
| Distribution | US Nationwide distribution - CA, IL, NY, PA, TN, TX |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported May 24, 2017. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.