Aptio Automation Storage and Retrieval Module (SRM)

Reported: July 12, 2023 Initiated: May 16, 2023 #Z-2091-2023

Product Description

Aptio Automation Storage and Retrieval Module (SRM)

Reason for Recall

A firmware error may lead to an incorrect association of test result(s) to a sample ID and therefore to an incorrect patient result. The mis-association of test results, or a delay in testing, may occur when all the conditions listed below occur within a few milliseconds timeframe of each other: - The SRM module is releasing a sample tube (Tube A) just placed into the carrier - Another sample tube (Tube B) is erroneously not diverted into the SRM module buffer - There is divert gate malfunction and a narrow timing window Only in this specific scenario, Tube A may be released by the module as Tube B due to a miscommunication between the module firmware and the Automation software without a unique error message being generated. The Automation System loses the traceability of Tube A. It manages both Tube A (incorrectly identified as Tube B) and the real Tube B according to the pending test orders (not yet performed) on Tube B.

Details

Units Affected
83 units
Distribution
Worldwide distribution - US Nationwide distribution in the states of CA, CO, FL, IL, KY, LA, MA, MD, MI, MS, MT, NC, NH, NY, OH, PA, SC, TN, TX, VA, WA, & WV. The countries of Australia, Brazil, Canada, China, France, Germany, Greece, Italy, Portugal, Saudi Arabia, Spain, Sweden, & United Kingdom.
Location
Newark, DE

Frequently Asked Questions

What product was recalled?
Aptio Automation Storage and Retrieval Module (SRM). Recalled by Siemens Healthcare Diagnostics, Inc.. Units affected: 83 units.
Why was this product recalled?
A firmware error may lead to an incorrect association of test result(s) to a sample ID and therefore to an incorrect patient result. The mis-association of test results, or a delay in testing, may occur when all the conditions listed below occur within a few milliseconds timeframe of each other: - The SRM module is releasing a sample tube (Tube A) just placed into the carrier - Another sample tube (Tube B) is erroneously not diverted into the SRM module buffer - There is divert gate malfunction and a narrow timing window Only in this specific scenario, Tube A may be released by the module as Tube B due to a miscommunication between the module firmware and the Automation software without a unique error message being generated. The Automation System loses the traceability of Tube A. It manages both Tube A (incorrectly identified as Tube B) and the real Tube B according to the pending test orders (not yet performed) on Tube B.
Which agency issued this recall?
This recall was issued by the FDA Devices on July 12, 2023. Severity: Moderate. Recall number: Z-2091-2023.