PlainRecalls
FDA Devices Moderate Class II Ongoing

BD Synapsys Informatics Solution, Catalog Number 444150

Reported: July 12, 2023 Initiated: June 8, 2023 #Z-2099-2023

Product Description

BD Synapsys Informatics Solution, Catalog Number 444150

Reason for Recall

There is a risk of mis-association of patient demographic data for users who choose to report results manually when their Service Control Unit (SCU) is operating near capacity and are using save actions (Skip, Final Discard, Mark as Read, Save) to process patient results rapidly.

Details

Recalling Firm
Becton Dickinson & Co.
Units Affected
102 systems
Distribution
US States: CA, CO, IA, IL, MI, MN, MO, NY, OH, OK, SC, TN, TX, UT, WI. Australia, Austria, Belgium, Canada, Denmark, France, Germany, Italy, Japan, Luxembourg, Netherlands, New Zealand, Norway, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, United Kingdom
Location
Sparks, MD

Frequently Asked Questions

What product was recalled?
BD Synapsys Informatics Solution, Catalog Number 444150. Recalled by Becton Dickinson & Co.. Units affected: 102 systems.
Why was this product recalled?
There is a risk of mis-association of patient demographic data for users who choose to report results manually when their Service Control Unit (SCU) is operating near capacity and are using save actions (Skip, Final Discard, Mark as Read, Save) to process patient results rapidly.
Which agency issued this recall?
This recall was issued by the FDA Devices on July 12, 2023. Severity: Moderate. Recall number: Z-2099-2023.