Severity
Moderate
FDA Devices recall · Reported June 13, 2018
The firm will be updating the Instructions for Use for the product. Current IFUs instruct users to lower the primary bag 20 inches below the secondary bag using a hanger when perf…
Baxter Healthcare Corporation recalled Baxter SIGMA Spectrum Infusion Pump with Master Drug Library (Version 8), infusion pump, … - a moderate-severity action.
Baxter SIGMA Spectrum Infusion Pump with Master Drug Library (Version 8), infusion pump, … was recalled by Baxter Healthcare Corporation in June 13, 2018. Reason: The firm will be updating the Instructions for Use for the product. Current IFUs instruct users to lower the …. Check the official notice for the remedy. Verify recall #Z-2113-2018 with the FDA Devices before acting.
The recall
Baxter Healthcare Corporation issued this moderate-severity FDA Devices recall-The firm will be updating the Instructions for Use for the product. Current IFUs instruct users to lower the ….
Sourced from official FDA Devices enforcement records. Verify recall #Z-2113-2018 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-2113-2018) was formally reported on June 13, 2018, with the manufacturer initiating the action on May 8, 2018. It is classified under Moderate severity (Class II), with a current status of Terminated. Baxter Healthcare Corporation is listed as the recalling firm, operating out of Medina, NY. Federal records list the affected scope as 122889, a scale large enough to require multi-state distribution tracking.
The documented reason for this recall is: The firm will be updating the Instructions for Use for the product. Current IFUs instruct users to lower the primary bag 20 inches below the secondary bag using a hanger when performing a secondary infusion. However, th… Distribution data in the federal record shows the product reached: Worldwide Distribution - US Nationwide and the country of Canada.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
122889
Related Recalls
6
3 from same agency
Baxter SIGMA Spectrum Infusion Pump with Master Drug Library (Version 8), infusion pump, 35700BAX2 Product Usage: The Baxter SIGMA Spectrum Infusion Pump with Master Drug Library is intended to be used for the controlled administration of fluids. These fluids may include pharmaceutical drugs, blood, blood products and mixture of required patient therapy. The intended routes of administration consist of the following clinically accepted routes: intravenous, arterial, subcutaneous, intrathecal, epidural or irrigation of fluid space. Spectrum Infusion Pump is intended to be used in conjunction with legally marketed intravenous administration sets and medications provided by the user. The Spectrum and Spectrum with Master Drug Library is suitable for many user facility applications such as but not limited to hospitals, outpatient care areas, homecare and ambulatory care services. Spectrum and Spectrum with Master Drug Library is intended to reduce operator interaction through automated programming thereby helping to reduce errors associated with complex device programming. Parameter programming requires trained healthcare professional confirmation of limits and drug therapy to the physicians directive.
The firm will be updating the Instructions for Use for the product. Current IFUs instruct users to lower the primary bag 20 inches below the secondary bag using a hanger when performing a secondary infusion. However, the firm supplies 12.5-inch hangers with the Baxter disposable secondary sets.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-2113-2018 |
| Date reported | June 13, 2018 |
| Date initiated | May 8, 2018 |
| Recalling firm | Baxter Healthcare Corporation |
| Firm location | Medina, NY |
| Affected scope | 122889 |
| Distribution | Worldwide Distribution - US Nationwide and the country of Canada. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported June 13, 2018. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.