Severity
Moderate
FDA Devices recall · Reported September 4, 2013
The ADVIA 120 optics cover or hood is located on the very top of the ADVIA 120 analyzer. When the optics cover is raised, it is supported by two gas struts (SMN 10309266) attache…
Siemens Healthcare Diagnostics recalled ADVIA 120 Hematology Analyzer, Catalog Numbers SMN 10360958 and 10360959 . The ADVIA 1… - a moderate-severity action.
ADVIA 120 Hematology Analyzer, Catalog Numbers SMN 10360958 and 10360959 . The ADVIA 1… was recalled by Siemens Healthcare Diagnostics in September 4, 2013. Reason: The ADVIA 120 optics cover or hood is located on the very top of the ADVIA 120 analyzer. When the optics cov…. Check the official notice for the remedy. Verify recall #Z-2114-2013 with the FDA Devices before acting.
The recall
Siemens Healthcare Diagnostics issued this moderate-severity FDA Devices recall-The ADVIA 120 optics cover or hood is located on the very top of the ADVIA 120 analyzer. When the optics cov….
Sourced from official FDA Devices enforcement records. Verify recall #Z-2114-2013 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-2114-2013) was formally reported on September 4, 2013, with the manufacturer initiating the action on June 26, 2013. It is classified under Moderate severity (Class II), with a current status of Terminated. Siemens Healthcare Diagnostics is listed as the recalling firm, operating out of Tarrytown, NY. Federal records list the affected scope as Domestic: 950 units; Foreign: 3369 units.
The documented reason for this recall is: The ADVIA 120 optics cover or hood is located on the very top of the ADVIA 120 analyzer. When the optics cover is raised, it is supported by two gas struts (SMN 10309266) attached to the cover on the left and right sid… Distribution data in the federal record shows the product reached: Worldwide Distribution - USA (nationwide) and the countries of Austria, Australia, Belgium, Bosnia - Herzegovina, China, Croatia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hungary, Iceland, Indi…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
Domestic: 950 units; Foreign: 3369 units
Related Recalls
6
3 from same agency
ADVIA 120 Hematology Analyzer, Catalog Numbers SMN 10360958 and 10360959 . The ADVIA 120 is a hematology system that utilizes the principles of flow cytometry in order to provide complete blood counts.
The ADVIA 120 optics cover or hood is located on the very top of the ADVIA 120 analyzer. When the optics cover is raised, it is supported by two gas struts (SMN 10309266) attached to the cover on the left and right sides. Over time, it is possible that these struts may lose their effectiveness and fail to support the optics cover in the open position. This could cause the cover to fall during maintenance procedures.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-2114-2013 |
| Date reported | September 4, 2013 |
| Date initiated | June 26, 2013 |
| Recalling firm | Siemens Healthcare Diagnostics |
| Firm location | Tarrytown, NY |
| Affected scope | Domestic: 950 units; Foreign: 3369 units |
| Distribution | Worldwide Distribution - USA (nationwide) and the countries of Austria, Australia, Belgium, Bosnia - Herzegovina, China, Croatia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hungary, Iceland, India, Indonesia, Irelan… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported September 4, 2013. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.