Equinoxe Reverse Shoulder Fracture Adapter Tray Screw Intended for Primary and Reverse Total Shoulder Arthroplasty.
Reported: August 8, 2012 Initiated: June 15, 2012 #Z-2118-2012
Product Description
Equinoxe Reverse Shoulder Fracture Adapter Tray Screw Intended for Primary and Reverse Total Shoulder Arthroplasty.
Reason for Recall
Exactech, Inc. initiated a recall of Equinoxe Replicator Plates, Equinoxe Locking Screws, Equinoxe Compression Screw Kits, Equinoxe Fracture Adapter Tray Screws, and Octane-C Intervertebral fusion devices due to the firm determining that the device labeling identifies the affected parts as having a ten year shelf life; however, the packaging configuration that was used it qualified to a five year
Details
- Recalling Firm
- Exactech, Inc.
- Units Affected
- 54,877 units for all products in Recall Event
- Distribution
- Worldwide Distribution - USA including FL, MO, MA, AL, KY, IN, NC, LA,PA, VA, OH, NY, ME, MI, TX, KS, TN, OK, NJ, WA, CO, GA, CA, MN, IA, OR, and IL and Internationaly to Austria, Luxenbourg, Sweden, Germany, Spain, France, Greece, Italy, Nederland, United Kingdom, Canada, India, Australia, Switzerland, and Sao Palo.
- Location
- Gainesville, FL
Frequently Asked Questions
What product was recalled? ▼
Equinoxe Reverse Shoulder Fracture Adapter Tray Screw Intended for Primary and Reverse Total Shoulder Arthroplasty.. Recalled by Exactech, Inc.. Units affected: 54,877 units for all products in Recall Event.
Why was this product recalled? ▼
Exactech, Inc. initiated a recall of Equinoxe Replicator Plates, Equinoxe Locking Screws, Equinoxe Compression Screw Kits, Equinoxe Fracture Adapter Tray Screws, and Octane-C Intervertebral fusion devices due to the firm determining that the device labeling identifies the affected parts as having a ten year shelf life; however, the packaging configuration that was used it qualified to a five year
Which agency issued this recall? ▼
This recall was issued by the FDA Devices on August 8, 2012. Severity: Moderate. Recall number: Z-2118-2012.
Related Recalls
FDA Devices Moderate
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
FDA Devices Moderate
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
FDA Devices Moderate
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11