FDA Devices recall · Reported August 6, 2014
Medtronic Nexdrive Micropositioning Drive. Models Ml-1000 and Ml-2000. For use in conjunction with the Medtronic Nexframe Stereotactic System for the precise positioning of microelectrodes and implantable leads. A stereotactic guidance system used in conjunction with Medtronic StealthStation Navigation Systems-image-guided surgery (IGS) systems-for Deep Brain Stimulation (DBS) procedures.
Potential for misalignment of the Z-stage scale. Using one of these devices for a procedure could result in the microelectrode being inserted to an incorrect target depth.
Medtronic Neuromodulation recalled Medtronic Nexdrive Micropositioning Drive. Models Ml-1000 and Ml-2000. For use in co… - a moderate-severity action.
Medtronic Nexdrive Micropositioning Drive. Models Ml-1000 and Ml-2000. For use in co… was recalled by Medtronic Neuromodulation in August 6, 2014. Reason: Potential for misalignment of the Z-stage scale. Using one of these devices for a procedure could result in t…. Check the official notice for the remedy. Verify recall #Z-2118-2014 with the FDA Devices before acting.
The recall
Medtronic Neuromodulation issued this moderate-severity FDA Devices recall-Potential for misalignment of the Z-stage scale. Using one of these devices for a procedure could result in t….
- Moderate
- severity level
- Class II
- classification
- August 6, 2014
- reported
Sourced from official FDA Devices enforcement records. Verify recall #Z-2118-2014 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
Recall Insight
This recall's affected scope, documented in the Recall Profile below, stays within a single-region or limited-distribution footprint. Distribution scope (see Distribution below) directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Medical Devices recalls over time
Where this recall sits in its category - 41,021 medical devices recalls on record
Of 88,244 recalls with an agency hazard classification, 19,068 are high severity, 64,652 moderate, and 4,524 low. A further 16,179 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Product description
Medtronic Nexdrive Micropositioning Drive. Models Ml-1000 and Ml-2000. For use in conjunction with the Medtronic Nexframe Stereotactic System for the precise positioning of microelectrodes and implantable leads. A stereotactic guidance system used in conjunction with Medtronic StealthStation Navigation Systems-image-guided surgery (IGS) systems-for Deep Brain Stimulation (DBS) procedures.
Reason for recall
Potential for misalignment of the Z-stage scale. Using one of these devices for a procedure could result in the microelectrode being inserted to an incorrect target depth.
Recall Profile
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-2118-2014 |
| Date reported | August 6, 2014 |
| Date initiated | June 11, 2014 |
| Recalling firm | Medtronic Neuromodulation |
| Firm location | Minneapolis, MN |
| Affected scope | 300 EXPANDED 10/02/2014 130 devices |
| Distribution | Worldwide Distribution - USA including CA, CO, FL, IL, KS, KY, LA, MN, MT, NC, NE, OR, TX, VA, and UT, and Internationally to Canada, Spain, and United Kingdom. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
- Check the recall number (Z-2118-2014) and product description against the item you own. Search the archive
- Confirm the current status with FDA Devices before acting, recall details can be updated.
- Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Where was the recalled product distributed? ▼
Should I stop using a recalled medication or medical device? ▼
Nationwide recalls with similar severity or category timing
Drug and device notices outside this firm and state, matched by severity timing and product category.
Similar severity · nearest report date
Same archive severity grade (Moderate), closest by report date among other firms and states.
Similar Medical Devices · nearest report date
Same product category, closest by report date among other firms and states (deduped against the severity set).
Compare this recall with GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VS… → · All Medical Devices recalls
Methodology · Corrections · Archive as of August 2026