Microscan rapID/S panels NC3.12 Sold in Japan only. Product Usage: MicroScan¿ Synergies plus¿ Panels & MicroScan¿ Rapids plus (Negative) - For use in determining quantitative and/or qualitative antimicrobial agent susceptibility and/or identification for gram-negative bacteria.

Confirmed increase in false susceptible and false intermediate misreads for Imipenem and/or meropenum on Synergies plus Negative and rap ID/S plus Negative panels read on Walk Awa…

Recall #
Z-2128-2013
Affected scope
57,600 panels in total
Initiated
August 21, 2013
Compiled from official public sources by the editorial team.
Verify with FDA Devices →
View my saved recalls

Siemens Healthcare Diagnostics, Inc. recalled Microscan rapID/S panels NC3.12 Sold in Japan only. Product Usage: MicroScan¿ Syn… - a critical-severity action.

Microscan rapID/S panels NC3.12 Sold in Japan only. Product Usage: MicroScan¿ Syn… was recalled by Siemens Healthcare Diagnostics, Inc. in September 18, 2013. Reason: Confirmed increase in false susceptible and false intermediate misreads for Imipenem and/or meropenum on Syne…. Check the official notice for the remedy. Verify recall #Z-2128-2013 with the FDA Devices before acting.

The recall

Siemens Healthcare Diagnostics, Inc. issued this critical-severity FDA Devices recall-Confirmed increase in false susceptible and false intermediate misreads for Imipenem and/or meropenum on Syne….

Critical
severity level
Class I
classification
September 18, 2013
reported

Sourced from official FDA Devices enforcement records. Verify recall #Z-2128-2013 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.

Recall Insight

This FDA Devices action (record #Z-2128-2013) was formally reported on September 18, 2013, with the manufacturer initiating the action on August 21, 2013. It is classified under Critical severity (Class I), with a current status of Terminated. Siemens Healthcare Diagnostics, Inc. is listed as the recalling firm, operating out of West Sacramento, CA. Federal records list the affected scope as 57,600 panels in total.

The documented reason for this recall is: Confirmed increase in false susceptible and false intermediate misreads for Imipenem and/or meropenum on Synergies plus Negative and rap ID/S plus Negative panels read on Walk Away System. Distribution data in the federal record shows the product reached: Worldwide Distribution - USA Nationwide in the states of: AK, SC, IA, MO GA, WI, TX, KS, CA, IL, CO, AZ, SD, IN, MD, FL, MN and the countries of:Japan, France, Algeria, Germany, Austria, Romania, Russian Federation, So…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.

Medical Devices recalls over time

Where this recall sits in its category - 40,322 medical devices recalls on record

-1,00001,0002,0003,0004,000 20052008201120142017202020232026 1,248
Severity2487872426High severity (most serious)Moderate severityLow severity
Where this recall sits in the database

Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified high severity.

Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.

Severity

Critical

Affected scope

57,600 panels in total

Related Recalls

6

3 from same agency

Product description

Microscan rapID/S panels NC3.12 Sold in Japan only. Product Usage: MicroScan¿ Synergies plus¿ Panels & MicroScan¿ Rapids plus (Negative) - For use in determining quantitative and/or qualitative antimicrobial agent susceptibility and/or identification for gram-negative bacteria.

Reason for recall

Confirmed increase in false susceptible and false intermediate misreads for Imipenem and/or meropenum on Synergies plus Negative and rap ID/S plus Negative panels read on Walk Away System.

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Critical (Class I)
Status Terminated
Recall number Z-2128-2013
Date reported September 18, 2013
Date initiated August 21, 2013
Recalling firm Siemens Healthcare Diagnostics, Inc.
Firm location West Sacramento, CA
Affected scope 57,600 panels in total
Distribution Worldwide Distribution - USA Nationwide in the states of: AK, SC, IA, MO GA, WI, TX, KS, CA, IL, CO, AZ, SD, IN, MD, FL, MN and the countries of:Japan, France, Algeria, Germany, Austria, Romania, Russian Federation, South Africa, Kazakhst…

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

What to do with this recall

Match your product against the recall record, then act on the agency remedy.

  • Check the recall number (Z-2128-2013) and product description against the item you own. Search the archive
  • Confirm the current status with FDA Devices before acting, recall details can be updated.
  • This is a Class I (high-risk) recall, stop using the product immediately and follow the disposal or return instructions. What to do next

This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.

Frequently Asked Questions

What product was recalled?
Microscan rapID/S panels NC3.12 Sold in Japan only. Product Usage: MicroScan¿ Synergies plus¿ Panels & MicroScan¿ Rapids plus (Negative) - For use in determining quantitative and/or qualitative antimicrobial agent susceptibility and/or identification for gram-negative bacteria.. Recalled by Siemens Healthcare Diagnostics, Inc.. Units affected: 57,600 panels in total.
Why was this product recalled?
Confirmed increase in false susceptible and false intermediate misreads for Imipenem and/or meropenum on Synergies plus Negative and rap ID/S plus Negative panels read on Walk Away System.
Which agency issued this recall?
This recall was issued by the FDA Devices on September 18, 2013. Severity: Critical. Recall number: Z-2128-2013.
Where was the recalled product distributed?
Distribution: Worldwide Distribution - USA Nationwide in the states of: AK, SC, IA, MO GA, WI, TX, KS, CA, IL, CO, AZ, SD, IN, MD, FL, MN and the countries of:Japan, France, Algeria, Germany, Austria, Romania, Russian Federation, South Africa, Kazakhstan, Belgium, Great Britain, Turkey, Spain, Italy, Saudi Arabia, Qutar, Poland, Iraq, Unite Arab Emrites, Algeria, Oman, Greece, Portugal, and Ireland..
How do I check if my product is affected by a recall?
Check the product description and recall number (Z-2128-2013) against your product. Visit the official FDA Devices website for the most current information. You can also use our Recall Checker tool to search by product name or brand.
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Nearby Recalls in This Category

Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.

Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →

Browse all Medical Devices recalls →

Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported September 18, 2013. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.