PlainRecalls
FDA Devices Moderate Class II Terminated

Cios Spin Image-Intensified Fluoroscopic X-Ray System, Model Numbers VA30-10308194

Reported: June 3, 2020 Initiated: April 17, 2020 #Z-2128-2020

Product Description

Cios Spin Image-Intensified Fluoroscopic X-Ray System, Model Numbers VA30-10308194

Reason for Recall

The unplugging of the main cable on the monitor trolley from the X10 connector can cause an unsafe electrical voltage (up to 67,6 V DC) at the X10 connector. This can result in an electrical shock if the user or other persons touch the contact pins of the X10 connector.

Details

Units Affected
20
Distribution
US Nationwide Distribution to states of: AR, AZ, CA, FL, IL, IN, MA, MD, MI, MN, NC, NE, NH, NJ, NY, OH, OK, SC, TN, TX, UT, VA, WA, and Puerto Rico.
Location
Malvern, PA

Frequently Asked Questions

What product was recalled?
Cios Spin Image-Intensified Fluoroscopic X-Ray System, Model Numbers VA30-10308194. Recalled by Siemens Medical Solutions USA, Inc. Units affected: 20.
Why was this product recalled?
The unplugging of the main cable on the monitor trolley from the X10 connector can cause an unsafe electrical voltage (up to 67,6 V DC) at the X10 connector. This can result in an electrical shock if the user or other persons touch the contact pins of the X10 connector.
Which agency issued this recall?
This recall was issued by the FDA Devices on June 3, 2020. Severity: Moderate. Recall number: Z-2128-2020.