FDA Devices Verify with FDA Devices → Moderate Class II Terminated

GE Healthcare, Discovery MR750w

Reported: July 13, 2016 Initiated: June 8, 2016 #Z-2131-2016 471 (184 US; 287 OUS) units

GE Healthcare, LLC issued this FDA Devices recall on July 13, 2016. Classified as Moderate severity (Class II). Approximately 471 (184 US; 287 OUS) units are affected. The recall was issued because: GE Healthcare has recently become aware of a potential safety issue with the patient bore heating on the Discovery MR75…. This recall notice is sourced from official FDA Devices enforcement records. Below you will find the complete product description, hazard information, remedy instructions, and related recalls from the same manufacturer or product category.

Recall Insight

This FDA Devices action (record #Z-2131-2016) was formally reported on July 13, 2016, with the manufacturer initiating the action on June 8, 2016. It is classified under Moderate severity (Class II), with a current status of Terminated. GE Healthcare, LLC is listed as the recalling firm, operating out of Waukesha, WI. Federal records indicate 471 (184 US; 287 OUS) units are affected, placing this recall in the million-unit bracket that typically triggers nationwide consumer alerts and retailer sweeps.

The documented reason for this recall is: GE Healthcare has recently become aware of a potential safety issue with the patient bore heating on the Discovery MR750w. A small area on the bore surface, which is normally accessible to the patient, can get warm e… Distribution data in the federal record shows the product reached: US: AK, AZ, CA, COM, CT , DC, FL, GA, GU, HI, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MO, NV, NH, NJ, NY, NC, OH,OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WY. OUS: Algeria Argentina Australia Austria Bahrain…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

To put this record in context, PlainRecalls indexes 83,949 recalls across the FDA, CPSC, NHTSA and USDA FSIS going back to 1995. Within the same product category, the database holds 6 closely related recalls, of which 6 were also issued by FDA Devices. That clustering is a signal — repeated actions in a narrow category often indicate a systemic quality-control issue, a supplier-wide contamination, or a design defect that has propagated across product lines. This recall is roughly 10 years old; older recalls can remain relevant because many units enter resale, rental, and secondary-market channels where the original warning never reaches the end user. Always cross-check the recall number against the official agency page before relying on any summary.

Recall Distribution by Severity Class

Severity1Class I (Critical)Class II (Moderate)Class III (Low)
Recall Distribution by Severity Class

Severity

Moderate

Units Affected

471 (184 US; 287 OUS)

Related Recalls

6

6 from same agency

Product Description

GE Healthcare, Discovery MR750w

Reason for Recall

GE Healthcare has recently become aware of a potential safety issue with the patient bore heating on the Discovery MR750w. A small area on the bore surface, which is normally accessible to the patient, can get warm enough to cause a serious burn in the event the patient touches the bore and proper padding is not used. No injury has been reported to GE as a result of this issue.

Details

Recalling Firm
GE Healthcare, LLC
Units Affected
471 (184 US; 287 OUS)
Distribution
US: AK, AZ, CA, COM, CT , DC, FL, GA, GU, HI, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MO, NV, NH, NJ, NY, NC, OH,OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WY. OUS: Algeria Argentina Australia Austria Bahrain Belarus Belgium Bermuda Brazil Bulgaria Canada China Egypt France Germany Hungary India Indonesia Ireland Italy Japan Kazakhstan Korea (Republic of) Kuwait Lebanon Mexico Netherlands New Zealand Poland Qatar Romania Russia Russian Federation Saudi Arabia Singapore Spain Sweden Switzerland Taiwan Thailand Turkey United Arab Emirates United Kingdom Uruguay
Location
Waukesha, WI

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Terminated
Recall number Z-2131-2016
Date reported July 13, 2016
Date initiated June 8, 2016
Recalling firm GE Healthcare, LLC
Units affected 471 (184 US; 287 OUS)
Distribution US: AK, AZ, CA, COM, CT , DC, FL, GA, GU, HI, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MO, NV, NH, NJ, NY, NC, OH,OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WY. OUS: Algeria Argentina Australia Austria Bahrain Belarus Belgium …

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

Scale of Impact

471 (184 US; 287 OUS) units affected — million-unit bracket.

Regional (<10K units)
Multi-state (10K – 100K units)
Large-scale (100K – 1M units)
Massive (≥1M units) ✓ This recall

Bracket cutoffs follow federal recall-disclosure conventions; bar widths scale linearly within each bracket. Source: PlainRecalls analysis of U.S. Food and Drug Administration filings.

Frequently Asked Questions

What product was recalled?
GE Healthcare, Discovery MR750w. Recalled by GE Healthcare, LLC. Units affected: 471 (184 US; 287 OUS).
Why was this product recalled?
GE Healthcare has recently become aware of a potential safety issue with the patient bore heating on the Discovery MR750w. A small area on the bore surface, which is normally accessible to the patient, can get warm enough to cause a serious burn in the event the patient touches the bore and proper padding is not used. No injury has been reported to GE as a result of this issue.
Which agency issued this recall?
This recall was issued by the FDA Devices on July 13, 2016. Severity: Moderate. Recall number: Z-2131-2016.
Where was the recalled product distributed?
Distribution: US: AK, AZ, CA, COM, CT , DC, FL, GA, GU, HI, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MO, NV, NH, NJ, NY, NC, OH,OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WY. OUS: Algeria Argentina Australia Austria Bahrain Belarus Belgium Bermuda Brazil Bulgaria Canada China Egypt France Germany Hungary India Indonesia Ireland Italy Japan Kazakhstan Korea (Republic of) Kuwait Lebanon Mexico Netherlands New Zealand Poland Qatar Romania Russia Russian Federation Saudi Arabia Singapore Spain Sweden Switzerland Taiwan Thailand Turkey United Arab Emirates United Kingdom Uruguay.
How do I check if my product is affected by a recall?
Check the product description and recall number (Z-2131-2016) against your product. Visit the official FDA Devices website for the most current information. You can also use our Recall Checker tool to search by product name or brand.
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Recall Context

Product recalls are issued when a manufacturer, distributor, or federal agency determines that a product poses a safety risk to consumers. This recall is classified as moderate severity, indicating the product may cause temporary or medically reversible health consequences. Across PlainRecalls, we track 83,000+ recalls from FDA, CPSC, and NHTSA to help consumers stay informed and act quickly when safety issues arise.

Nearby Recalls in This Category

Other recalls in the same product category — useful for spotting patterns across the same defect class or manufacturer.

Compare this recall with Endo-Model Replacement Plateau; Item Number: 15-0027/11; →

Data Sources

Data as of 2025. Source: FDA, CPSC, NHTSA, USDA FSIS federal recall databases.

  • Source: FDA — Food and Drug Administration, openFDA Enforcement API (food, drug, and medical device recalls)
  • Source: CPSC — Consumer Product Safety Commission Recalls API (consumer product recalls and hazards)
  • Source: NHTSA — National Highway Traffic Safety Administration Recalls API (vehicle safety recalls)
  • Source: USDA FSIS — Food Safety and Inspection Service (meat, poultry, and egg product recalls)

Recall information is sourced from official federal agency databases. Always verify recall details with the issuing agency for the most current status. This information is for research and awareness purposes only.

All federal data sources used on this page

Source: Federal recall agencies (FDA, CPSC, NHTSA, USDA FSIS) Aggregated multi-agency recall feeds · 2024 Recall data normalized across federal agency feeds; severity classifications follow each agency's own taxonomy (FDA Class I/II/III; CPSC, NHTSA, USDA FSIS).