RF Cannula, 100 mm,Curved 5 MM, 20 G, REF 0406-630-115 RX only, Disposable ( 10/pk) Product Usage: The Stryker RF Cannulae, in combination with the Stryker RF Generator/Multi- Gen and electrodes, are intended for coagulation of soft tissue in orthopedic, arthroscopic, spinal, and neurosurgical applications.

Reported: August 8, 2012 Initiated: July 17, 2012 #Z-2134-2012

Product Description

RF Cannula, 100 mm,Curved 5 MM, 20 G, REF 0406-630-115 RX only, Disposable ( 10/pk) Product Usage: The Stryker RF Cannulae, in combination with the Stryker RF Generator/Multi- Gen and electrodes, are intended for coagulation of soft tissue in orthopedic, arthroscopic, spinal, and neurosurgical applications.

Reason for Recall

Stryker has received 6 complaints that the 5mm needle was placed in 10 mm package and 10 mm needle was placed in the 5 mm package. If the clinician does not recognize that the active tip on the cannula is 10 mm instead of 5mm and the cannula is used in a cervical procedure there is a potential for a thermal skin lesion due to a larger lesion creation. If the 10 mm active tip cannula, as opposed

Details

Units Affected
210
Distribution
Worldwide Distribution - US Nationwide and the countries of England, France, Netherlands, and Switzerland.
Location
Portage, MI

Frequently Asked Questions

What product was recalled?
RF Cannula, 100 mm,Curved 5 MM, 20 G, REF 0406-630-115 RX only, Disposable ( 10/pk) Product Usage: The Stryker RF Cannulae, in combination with the Stryker RF Generator/Multi- Gen and electrodes, are intended for coagulation of soft tissue in orthopedic, arthroscopic, spinal, and neurosurgical applications.. Recalled by Stryker Instruments Div. of Stryker Corporation. Units affected: 210.
Why was this product recalled?
Stryker has received 6 complaints that the 5mm needle was placed in 10 mm package and 10 mm needle was placed in the 5 mm package. If the clinician does not recognize that the active tip on the cannula is 10 mm instead of 5mm and the cannula is used in a cervical procedure there is a potential for a thermal skin lesion due to a larger lesion creation. If the 10 mm active tip cannula, as opposed
Which agency issued this recall?
This recall was issued by the FDA Devices on August 8, 2012. Severity: Moderate. Recall number: Z-2134-2012.