Severity
Moderate
FDA Devices recall · Reported June 20, 2018
The syringe plunger on the device may be mounted in a tilted position, which applies mechanical stress to the plunger. In the worst case scenario, the plunger may crack leading to…
Roche Diagnostics Corporation recalled cobas e 801 immunoassay analyzer, cobas 8000 Modular Series The device is a fully auto… - a moderate-severity action.
cobas e 801 immunoassay analyzer, cobas 8000 Modular Series The device is a fully auto… was recalled by Roche Diagnostics Corporation in June 20, 2018. Reason: The syringe plunger on the device may be mounted in a tilted position, which applies mechanical stress to the…. Check the official notice for the remedy. Verify recall #Z-2136-2018 with the FDA Devices before acting.
The recall
Roche Diagnostics Corporation issued this moderate-severity FDA Devices recall-The syringe plunger on the device may be mounted in a tilted position, which applies mechanical stress to the….
Sourced from official FDA Devices enforcement records. Verify recall #Z-2136-2018 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-2136-2018) was formally reported on June 20, 2018, with the manufacturer initiating the action on April 3, 2018. It is classified under Moderate severity (Class II), with a current status of Terminated. Roche Diagnostics Corporation is listed as the recalling firm, operating out of Indianapolis, IN. Federal records list the affected scope as 35.
The documented reason for this recall is: The syringe plunger on the device may be mounted in a tilted position, which applies mechanical stress to the plunger. In the worst case scenario, the plunger may crack leading to impaired syringe function. Distribution data in the federal record shows the product reached: US Distribution to the states of : CA, CT, IA, IN, KY, ME, MI, NC, NY, OH, SC, TX, and VA.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
35
Related Recalls
6
3 from same agency
cobas e 801 immunoassay analyzer, cobas 8000 Modular Series The device is a fully automated, software controlled analyzer system for in vitro determination of analytes in human body fluids. It is part of the cobas 8000 modular analyzer series. It uses electrochemiluminescent technology for signal generation and measurement.
The syringe plunger on the device may be mounted in a tilted position, which applies mechanical stress to the plunger. In the worst case scenario, the plunger may crack leading to impaired syringe function.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-2136-2018 |
| Date reported | June 20, 2018 |
| Date initiated | April 3, 2018 |
| Recalling firm | Roche Diagnostics Corporation |
| Firm location | Indianapolis, IN |
| Affected scope | 35 |
| Distribution | US Distribution to the states of : CA, CT, IA, IN, KY, ME, MI, NC, NY, OH, SC, TX, and VA. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported June 20, 2018. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.