The PCA 3 system cam be used in a wide range of clinical settings, including but not limited to: general floor, labor/delivery,post-partum, burn unit, medical surgical, operating room, oncology, critical care units, post-anesthesia care unit (PACU), and pediatrics. The PCA 3 infusion pump allows clinicians to administer or patients to self-administer analgesia safely and effectively within clinician programmed limits. The epidural route can be used to provide anesthesia or analgesia.

Reported: July 20, 2016 Initiated: March 8, 2013 #Z-2140-2016

Product Description

The PCA 3 system cam be used in a wide range of clinical settings, including but not limited to: general floor, labor/delivery,post-partum, burn unit, medical surgical, operating room, oncology, critical care units, post-anesthesia care unit (PACU), and pediatrics. The PCA 3 infusion pump allows clinicians to administer or patients to self-administer analgesia safely and effectively within clinician programmed limits. The epidural route can be used to provide anesthesia or analgesia.

Reason for Recall

Hospira has received reports of PCA pumps not detecting distal occlusions, one of which resulted in a serious injury. The issue is caused by normal wear and tear on the Half Nut (the component/nut that travels up and down the lead screw) which prevents it from properly detecting the pressure build-up associated with a distal occlusion.

Details

Recalling Firm
Hospira Inc.
Units Affected
8,216 infusion pump systems
Distribution
Nationwide Distribution to AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY) + District of Columbia + Puerto Rico + Virgin Islands; *** FOREIGN:
Location
Lake Forest, IL

Frequently Asked Questions

What product was recalled?
The PCA 3 system cam be used in a wide range of clinical settings, including but not limited to: general floor, labor/delivery,post-partum, burn unit, medical surgical, operating room, oncology, critical care units, post-anesthesia care unit (PACU), and pediatrics. The PCA 3 infusion pump allows clinicians to administer or patients to self-administer analgesia safely and effectively within clinician programmed limits. The epidural route can be used to provide anesthesia or analgesia.. Recalled by Hospira Inc.. Units affected: 8,216 infusion pump systems.
Why was this product recalled?
Hospira has received reports of PCA pumps not detecting distal occlusions, one of which resulted in a serious injury. The issue is caused by normal wear and tear on the Half Nut (the component/nut that travels up and down the lead screw) which prevents it from properly detecting the pressure build-up associated with a distal occlusion.
Which agency issued this recall?
This recall was issued by the FDA Devices on July 20, 2016. Severity: Moderate. Recall number: Z-2140-2016.