Life2000 Ventilator, Product Code: MS-01-0118, which is contained in the Life2000 Ventilator System, with Product Codes BT-20-0002, BT-20-0002A and BT-20-0002AP. The Life2000 ventilation system includes the Life2000 ventilator which has its individual label.

Reported: July 3, 2024 Initiated: May 29, 2024 #Z-2140-2024

Product Description

Life2000 Ventilator, Product Code: MS-01-0118, which is contained in the Life2000 Ventilator System, with Product Codes BT-20-0002, BT-20-0002A and BT-20-0002AP. The Life2000 ventilation system includes the Life2000 ventilator which has its individual label.

Reason for Recall

Potential for certain Life2000 ventilator systems to either fail to charge or have intermittent charging behavior due to damage to the battery charger dongle. Damage of the battery charger dongle prevents the ventilator's internal battery from charging.

Details

Units Affected
2,510 units
Distribution
US Nationwide distribution.
Location
Deerfield, IL

Frequently Asked Questions

What product was recalled?
Life2000 Ventilator, Product Code: MS-01-0118, which is contained in the Life2000 Ventilator System, with Product Codes BT-20-0002, BT-20-0002A and BT-20-0002AP. The Life2000 ventilation system includes the Life2000 ventilator which has its individual label.. Recalled by Baxter Healthcare Corporation. Units affected: 2,510 units.
Why was this product recalled?
Potential for certain Life2000 ventilator systems to either fail to charge or have intermittent charging behavior due to damage to the battery charger dongle. Damage of the battery charger dongle prevents the ventilator's internal battery from charging.
Which agency issued this recall?
This recall was issued by the FDA Devices on July 3, 2024. Severity: Critical. Recall number: Z-2140-2024.