PlainRecalls
FDA Devices Moderate Class II Terminated

Maquet Cardiopulmonary (MCP)/Getinge HLS Set Advanced 7.0. Model Number BEQ-HLS-7050 USA, Part Number 701052794

Reported: September 25, 2019 Initiated: July 11, 2019 #Z-2149-2019

Product Description

Maquet Cardiopulmonary (MCP)/Getinge HLS Set Advanced 7.0. Model Number BEQ-HLS-7050 USA, Part Number 701052794

Reason for Recall

The sets are configured with quick connectors that have been assembled in reverse on the arterial and venous lines, resulting in blue (venous) to red (arterial) and red (arterial) to blue (venous) connections. Due to this incorrect assembly, the sets cannot be primed from the reservoir prior to use, and the device may not function as intended.

Details

Units Affected
54
Distribution
The products were distributed to the following US states: AR, CA, FL, GA, IA, IL, IN, KY, NC, NE, NM, NV, NY, OH, OR, TN, and WI.
Location
Wayne, NJ

Frequently Asked Questions

What product was recalled?
Maquet Cardiopulmonary (MCP)/Getinge HLS Set Advanced 7.0. Model Number BEQ-HLS-7050 USA, Part Number 701052794. Recalled by Maquet Cardiovascular Us Sales, Llc. Units affected: 54.
Why was this product recalled?
The sets are configured with quick connectors that have been assembled in reverse on the arterial and venous lines, resulting in blue (venous) to red (arterial) and red (arterial) to blue (venous) connections. Due to this incorrect assembly, the sets cannot be primed from the reservoir prior to use, and the device may not function as intended.
Which agency issued this recall?
This recall was issued by the FDA Devices on September 25, 2019. Severity: Moderate. Recall number: Z-2149-2019.