Centrella Max - The Centrella Smart+ Bed is intended for use in healthcare environments as a patient support system to prevent and/or treat pressure injuries.Multiple model numbers of Centrella Max mattresses and pro+ mattresses. (1) Centrella Max Surface, REF P7922A01; (2) Centrella Max Surface, REF P7922A02; (3) Centrella Max Surface with X-Ray, REF P7922A03; (4) Centrella Max,Narrow, Rental, REF P7922ARENT01; (5) Centrella Pro+ 36" Surface with X-Ray, REF P7923A01; (6) Centrella Pro+
Product Description
Centrella Max - The Centrella Smart+ Bed is intended for use in healthcare environments as a patient support system to prevent and/or treat pressure injuries.Multiple model numbers of Centrella Max mattresses and pro+ mattresses. (1) Centrella Max Surface, REF P7922A01; (2) Centrella Max Surface, REF P7922A02; (3) Centrella Max Surface with X-Ray, REF P7922A03; (4) Centrella Max,Narrow, Rental, REF P7922ARENT01; (5) Centrella Pro+ 36" Surface with X-Ray, REF P7923A01; (6) Centrella Pro+ 40" Surface with X-Ray, REF P7923A02; (7) Centrella Pro+36" Surface, REF P7923A03; (8) Centrella Pro+ 40" Surface, REF P7923A04; (9) Pro+ MRS Surface for Hillrom 900 with X-Ray, REF P006800A01; (10) 80" MRS with X-Ray AU/NZ REF P006800A03; (11) Pro+ MRS Surface for Versacare with X-Ray, REF P3255A01; (12) Pro+ MRS Surface for Versacare, REF P3255A02; (13) Pro + MRS Surface for Versacare with X-Ray, REF P3255ARENT01; (14) Pro+ 36"MRS Surface with X-Ray, REF P7924A01; (15) Pro+ 40" MRS Surface with X-Ray, REF P7924A02; (16) Pro+ 36" MRS Surface, REF P7924A03; (17) Pro+ 40" MRS Surface, REF P7924A04; (18) Pro+ 36" MRS Rental Surface with X-Ray, REF P7924ARENT01.
Reason for Recall
There is a potential for the top cover of the hospital bed mattress to delaminate as a result of improper cleaning and lack of adequate details on the appropriate use of cleaning agents containing bleach.
Details
- Recalling Firm
- Baxter Healthcare Corporation
- Units Affected
- 38,136 mattress units and 20,187 service part units
- Distribution
- Distribution was made nationwide. There was also government/military distribution. Foreign distribution was made to Australia, Canada, China, Japan, Qatar, Saudi Arabia, Singapore, Taiwan, Turkey, and United Arab Emirates.
- Location
- Deerfield, IL
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11