Severity
Low
FDA Devices recall · Reported August 15, 2012
During the real-time stability testing for the cobas KRAS Mutation Test on the commercial lot PO6778, it was identified that the kit was just within specifications for the codon 6…
Roche Molecular Systems, Inc. recalled Cobas KRAS Mutation Test for In Vitro Diagnostic Use Product Usage: Usage: The Cobas… - a low-severity action.
Cobas KRAS Mutation Test for In Vitro Diagnostic Use Product Usage: Usage: The Cobas… was recalled by Roche Molecular Systems, Inc. in August 15, 2012. Reason: During the real-time stability testing for the cobas KRAS Mutation Test on the commercial lot PO6778, it was …. Check the official notice for the remedy. Verify recall #Z-2159-2012 with the FDA Devices before acting.
The recall
Roche Molecular Systems, Inc. issued this low-severity FDA Devices recall-During the real-time stability testing for the cobas KRAS Mutation Test on the commercial lot PO6778, it was ….
Sourced from official FDA Devices enforcement records. Verify recall #Z-2159-2012 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-2159-2012) was formally reported on August 15, 2012, with the manufacturer initiating the action on February 24, 2012. It is classified under Low severity (Class III), with a current status of Terminated. Roche Molecular Systems, Inc. is listed as the recalling firm, operating out of Branchburg, NJ. Federal records list the affected scope as 264 kits to 13 EU countries; 2 kits to Canada and 33 kits to 11 countries in the Rest of the World.
The documented reason for this recall is: During the real-time stability testing for the cobas KRAS Mutation Test on the commercial lot PO6778, it was identified that the kit was just within specifications for the codon 61 peak height ratio, and that the kit ha… Distribution data in the federal record shows the product reached: Worldwide Distribution - including the countries of Austria, Belgium, Czech Republic, Denmark, Germany, Italy, Luxemburg, Netherlands, Poland, Slovakia, Slovenia, Spain, United Kingdom, Canada, Chile, Greenland, Hong Ko…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified low severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Low
Affected scope
264 kits to 13 EU countries; 2 kits to Canada and 33 kits to 11 countries in the Rest of the World
Related Recalls
6
3 from same agency
Cobas KRAS Mutation Test for In Vitro Diagnostic Use Product Usage: Usage: The Cobas KRAS Mutation Test, for use with the Cobas 4800 System, is a real-time PCR test intended for the identification of mutations in codons 12, 13 and 61 of the KRAS gene in DNA derived from formalin-fixed paraffin-embedded human colorectal cancer (CRC).
During the real-time stability testing for the cobas KRAS Mutation Test on the commercial lot PO6778, it was identified that the kit was just within specifications for the codon 61 peak height ratio, and that the kit has trended towards the upper specification limit at the 6 month time point but did passed the 6 month stability testing. The kit is labeled for 9 months shelf life stability. The d
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Low (Class III) |
| Status | Terminated |
| Recall number | Z-2159-2012 |
| Date reported | August 15, 2012 |
| Date initiated | February 24, 2012 |
| Recalling firm | Roche Molecular Systems, Inc. |
| Firm location | Branchburg, NJ |
| Affected scope | 264 kits to 13 EU countries; 2 kits to Canada and 33 kits to 11 countries in the Rest of the World |
| Distribution | Worldwide Distribution - including the countries of Austria, Belgium, Czech Republic, Denmark, Germany, Italy, Luxemburg, Netherlands, Poland, Slovakia, Slovenia, Spain, United Kingdom, Canada, Chile, Greenland, Hong Kong, Israel, Malaysia… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported August 15, 2012. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.