Severity
Moderate
FDA Devices recall · Reported August 15, 2012
A customer has reported an error on an ABL8xx configured to a Hospital Information System using the data transfer protocol HL7 version 2.5. If the instrument transmits results t…
Radiometer America Inc recalled The ABL software version 6.12 for the ABL800 analyzer. The analyzers are intended for the… - a moderate-severity action.
The ABL software version 6.12 for the ABL800 analyzer. The analyzers are intended for the… was recalled by Radiometer America Inc in August 15, 2012. Reason: A customer has reported an error on an ABL8xx configured to a Hospital Information System using the data tran…. Check the official notice for the remedy. Verify recall #Z-2162-2012 with the FDA Devices before acting.
The recall
Radiometer America Inc issued this moderate-severity FDA Devices recall-A customer has reported an error on an ABL8xx configured to a Hospital Information System using the data tran….
Sourced from official FDA Devices enforcement records. Verify recall #Z-2162-2012 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-2162-2012) was formally reported on August 15, 2012, with the manufacturer initiating the action on July 5, 2012. It is classified under Moderate severity (Class II), with a current status of Terminated. Radiometer America Inc is listed as the recalling firm, operating out of Westlake, OH. Federal records list the affected scope as 11929 ABL 8xx units.
The documented reason for this recall is: A customer has reported an error on an ABL8xx configured to a Hospital Information System using the data transfer protocol HL7 version 2.5. If the instrument transmits results to a HIS system and the connection is bro… Distribution data in the federal record shows the product reached: Worldwide distribution: USA (nationwide) including the states of AK, AL, AR, AZ, BM, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
11929 ABL 8xx units
Related Recalls
6
3 from same agency
The ABL software version 6.12 for the ABL800 analyzer. The analyzers are intended for the in Vitro testing of: -Samples of whole blood for the parameters pH, pO2, pCO2, cK+, cNa+, cCa2+, cCl, cGlu, cLac, cCrea, ctBil, and co-oximetry parameters (ctHb, sO2, and the hemoglobin fractions FO2Hb, FCOHb, FMetHb, FHHb and FHbF) -Samples of expired air for the parameters pO2 and pCO2 -Pleura samples for the pH parameter. The ABL software version 6.12 for the ABL800 and ABL700 analyzer. The analyzers are intended for the in Vitro testing of: -Samples of whole blood for the parameters pH, pO2, pCO2, cK+, cNa+, cCa2+, cCl-, cGlu, cLac, cCrea, ctBil, and co-oximetry parameters (ctHb, sO2, and the hemoglobin fractions FO2Hb, FCOHb, FMetHb, FHHb and FHbF) -Samples of expired air for the parameters pO2 and pCO2 -Pleura samples for the pH parameter.
A customer has reported an error on an ABL8xx configured to a Hospital Information System using the data transfer protocol HL7 version 2.5. If the instrument transmits results to a HIS system and the connection is broken the results are queued for later transmission. If the ABL and HIS system are configured with HL7 2.5, any results in the queue will be mixed up when the instrument transmits t
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-2162-2012 |
| Date reported | August 15, 2012 |
| Date initiated | July 5, 2012 |
| Recalling firm | Radiometer America Inc |
| Firm location | Westlake, OH |
| Affected scope | 11929 ABL 8xx units |
| Distribution | Worldwide distribution: USA (nationwide) including the states of AK, AL, AR, AZ, BM, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported August 15, 2012. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.