PlainRecalls
ModerateClass IITerminated

FDA Devices recall · Reported August 20, 2014

EndoWrist Stapler System User Manual Addendum and Wall Chart; The EndoWrist Stapler 45 System and EndoWrist Stapler 45 Reloads are intended to be used with the da Vinci Si Surgical System (Model IS3000) for resection, transection and/or creation of anastomoses in General, Gynecologic and Urologic surgery. The device can be used with staple line or tissue buttressing material (natural or synthetic).

Changes were made to the Stapler EndoWrist Stapler 45 System Instruments and Accessories User Manual Addendum (550986-03) and the EndoWrist Stapler 45 Wall Chart (PN 551524-01) in…

Recall #
Z-2171-2014
Affected scope
2
Initiated
June 2, 2014
Verify with FDA Devices →
View my saved recalls

Intuitive Surgical, Inc. recalled EndoWrist Stapler System User Manual Addendum and Wall Chart; The EndoWrist Stapler 4… - a moderate-severity action.

EndoWrist Stapler System User Manual Addendum and Wall Chart; The EndoWrist Stapler 4… was recalled by Intuitive Surgical, Inc. in August 20, 2014. Reason: Changes were made to the Stapler EndoWrist Stapler 45 System Instruments and Accessories User Manual Addendum…. Check the official notice for the remedy. Verify recall #Z-2171-2014 with the FDA Devices before acting.

The recall

Intuitive Surgical, Inc. issued this moderate-severity FDA Devices recall-Changes were made to the Stapler EndoWrist Stapler 45 System Instruments and Accessories User Manual Addendum….

Moderate
severity level
2 units
affected scope
Class II
classification
August 20, 2014
reported

Sourced from official FDA Devices enforcement records. Verify recall #Z-2171-2014 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.

Recall Insight

This recall's affected scope, documented in the Recall Profile below, stays within a single-region or limited-distribution footprint. Distribution scope (see Distribution below) directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.

Medical Devices recalls over time

Where this recall sits in its category - 41,021 medical devices recalls on record

01,0002,0003,0004,000 20052008201120142017202020232026 1,947
Severity1906864652High severity (most serious)Moderate severityLow severity
Where this agency-classified recall sits among FDA Class I/II/III notices

Of 88,244 recalls with an agency hazard classification, 19,068 are high severity, 64,652 moderate, and 4,524 low. A further 16,179 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.

Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.

Product description

EndoWrist Stapler System User Manual Addendum and Wall Chart; The EndoWrist Stapler 45 System and EndoWrist Stapler 45 Reloads are intended to be used with the da Vinci Si Surgical System (Model IS3000) for resection, transection and/or creation of anastomoses in General, Gynecologic and Urologic surgery. The device can be used with staple line or tissue buttressing material (natural or synthetic).

Reason for recall

Changes were made to the Stapler EndoWrist Stapler 45 System Instruments and Accessories User Manual Addendum (550986-03) and the EndoWrist Stapler 45 Wall Chart (PN 551524-01) in which the pH range specification for the enzymatic cleaner used throughout the cleaning process was modified from pH-neutral to pH-neutral to mild alkaline (mild alkaline is defined as pHd11). Mild Alkaline enzymatic

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Terminated
Recall number Z-2171-2014
Date reported August 20, 2014
Date initiated June 2, 2014
Recalling firm Intuitive Surgical, Inc.
Firm location Sunnyvale, CA
Affected scope 2
Distribution Distribution to Florida and South Carolina

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

Scale of Impact

2 units affected - limited or regional distribution scale.

Regional (<10K units) ✓ This recall
Multi-state (10K – 100K units) -
Large-scale (100K – 1M units) -
Massive (≥1M units) -

Bracket cutoffs follow federal recall-disclosure conventions; bar widths scale linearly within each bracket. Source: PlainRecalls analysis of U.S. Food and Drug Administration filings.

What to do with this recall

Match your product against the recall record, then act on the agency remedy.

  • Check the recall number (Z-2171-2014) and product description against the item you own. Search the archive
  • Confirm the current status with FDA Devices before acting, recall details can be updated.
  • Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next

This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.

Frequently Asked Questions

What product was recalled?
EndoWrist Stapler System User Manual Addendum and Wall Chart; The EndoWrist Stapler 45 System and EndoWrist Stapler 45 Reloads are intended to be used with the da Vinci Si Surgical System (Model IS3000) for resection, transection and/or creation of anastomoses in General, Gynecologic and Urologic surgery. The device can be used with staple line or tissue buttressing material (natural or synthetic).. Recalled by Intuitive Surgical, Inc.. Units affected: 2.
Why was this product recalled?
Changes were made to the Stapler EndoWrist Stapler 45 System Instruments and Accessories User Manual Addendum (550986-03) and the EndoWrist Stapler 45 Wall Chart (PN 551524-01) in which the pH range specification for the enzymatic cleaner used throughout the cleaning process was modified from pH-neutral to pH-neutral to mild alkaline (mild alkaline is defined as pHd11). Mild Alkaline enzymatic
Which agency issued this recall?
This recall was issued by the FDA Devices on August 20, 2014. Severity: Moderate. Recall number: Z-2171-2014.
Where was the recalled product distributed?
Distribution: Distribution to Florida and South Carolina.
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Nationwide recalls with similar severity or category timing

Drug and device notices outside this firm and state, matched by severity timing and product category.

Compare this recall with GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VS… → · All Medical Devices recalls

This recall: FDA Devices, reported August 20, 2014. Severity classes follow each agency's own taxonomy; related-recall context is computed across the full archive.

Methodology · Corrections · Archive as of August 2026