Severity
Moderate
FDA Devices recall · Reported September 18, 2013
During an internal investigation, a formulation error that occurred was identified during the manufacture of the cobas¿ 4800 HPV Master Mix reagent that was packaged in four (4) c…
Roche Molecular Systems, Inc. recalled KIT cobas 4800 HPV AMP/DET 240T / 960T US-IVD, CE-IVD Roche Molecular System, Inc. 1080 … - a moderate-severity action.
KIT cobas 4800 HPV AMP/DET 240T / 960T US-IVD, CE-IVD Roche Molecular System, Inc. 1080 … was recalled by Roche Molecular Systems, Inc. in September 18, 2013. Reason: During an internal investigation, a formulation error that occurred was identified during the manufacture of …. Check the official notice for the remedy. Verify recall #Z-2173-2013 with the FDA Devices before acting.
The recall
Roche Molecular Systems, Inc. issued this moderate-severity FDA Devices recall-During an internal investigation, a formulation error that occurred was identified during the manufacture of ….
Sourced from official FDA Devices enforcement records. Verify recall #Z-2173-2013 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-2173-2013) was formally reported on September 18, 2013, with the manufacturer initiating the action on July 25, 2013. It is classified under Moderate severity (Class II), with a current status of Terminated. Roche Molecular Systems, Inc. is listed as the recalling firm, operating out of Branchburg, NJ. Federal records list the affected scope as US-IVD: 350 Kits; CE-IVD: 441 KITS.
The documented reason for this recall is: During an internal investigation, a formulation error that occurred was identified during the manufacture of the cobas¿ 4800 HPV Master Mix reagent that was packaged in four (4) cobas¿ 4800 AMP/DET kits, US- and CE-IVD … Distribution data in the federal record shows the product reached: Worldwide Distribution - USA (nationwide) including the countries of Austria, Australia, Belgium, Brazil, Czech Rep, Canada, Croatia, Columbia, Denmark, Dominican Republic, Germany, Ecuador, France, Japan, Greenland, Me…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
US-IVD: 350 Kits; CE-IVD: 441 KITS
Related Recalls
6
3 from same agency
KIT cobas 4800 HPV AMP/DET 240T / 960T US-IVD, CE-IVD Roche Molecular System, Inc. 1080 US highway 202 South Branchburg, NJ 08876 USA Roche Diagnostics 9115 Hague Road Indianapolis, IN 46250-0457 USA The cobas¿ HPV Test is a qualitative in vitro test for the detection of Human Papillomavirus in patient specimens. The test utilizes amplification of target DNA by the Polymerase Chain Reaction (PCR) and nucleic acid hybridization for the detection of 14 high-risk (HR) HPV types in a single analysis. The test specifically identifies types HPV16 and HPV18 while concurrently detecting the rest of the high risk types (31, 33, 35, 39, 45, 51, 52, 56, 58, 59, 66 and 68). The cobas¿ HPV Test is indicated: (a) To screen patients 21 years and older with ASC-US (Atypical squamous cells of undetermined significance) cervical cytology test results to determine the need for referral to colposcopy. (b) To be used in patients 21 years and older with ASC-US cervical cytology results, to assess the presence or absence of high-risk HPV genotypes 16 and 18. This information, together with the physicians assessment of cytology history, other risk factors, and professional guidelines, may be used to guide patient management. The results of this test are not intended to prevent women from proceeding to colposcopy. (c) In women 30 years and older, the cobas¿ HPV Test can be used with cervical cytology to adjunctively screen to assess the presence or absence of high risk HPV types. This information, together with the physicians assessment of cytology history, other risk factors, and professional guidelines, may be used to guide patient management. (d) In women 30 years and older, the cobas¿ HPV Test can be used to assess the presence or absence of HPV genotypes 16 and 18. This information, together with the physicians assessment of cytology history, other risk factors, and professional guidelines, may be used to guide patient management. Cervical specimens that may be tested with the cobas¿ HPV Test include the following liquid based collection media and collection device: " ThinPrep¿ Pap TestTM PreservCyt¿ Solution " Endocervical Brush/Spatula
During an internal investigation, a formulation error that occurred was identified during the manufacture of the cobas¿ 4800 HPV Master Mix reagent that was packaged in four (4) cobas¿ 4800 AMP/DET kits, US- and CE-IVD (240 and 960 test kit configurations). Examination of the batch production record of the material revealed there was an error in the tare weight used for the carboy during the fina
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-2173-2013 |
| Date reported | September 18, 2013 |
| Date initiated | July 25, 2013 |
| Recalling firm | Roche Molecular Systems, Inc. |
| Firm location | Branchburg, NJ |
| Affected scope | US-IVD: 350 Kits; CE-IVD: 441 KITS |
| Distribution | Worldwide Distribution - USA (nationwide) including the countries of Austria, Australia, Belgium, Brazil, Czech Rep, Canada, Croatia, Columbia, Denmark, Dominican Republic, Germany, Ecuador, France, Japan, Greenland, Mexico, Italy, New Zea… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported September 18, 2013. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.