PlainRecalls
FDA Devices Moderate Class II Ongoing

Dual Flat Panel Monitor (DFPM) accessories for HarmonyAIR E-Series Surgical Lighting Systems

Reported: August 6, 2025 Initiated: June 23, 2025 #Z-2178-2025

Product Description

Dual Flat Panel Monitor (DFPM) accessories for HarmonyAIR E-Series Surgical Lighting Systems

Reason for Recall

The DFPM yoke assembly ("yoke assembly") may detach from the spring arm. Should the yoke assembly detach from the spring arm, the yoke assembly would not be expected to fully detach but rather hang from the wiring within the yoke assembly.

Details

Recalling Firm
Steris Corporation
Units Affected
59 units
Distribution
Worldwide - US Nationwide and Puerto Rico and the country of Canada.
Location
Mentor, OH

Frequently Asked Questions

What product was recalled?
Dual Flat Panel Monitor (DFPM) accessories for HarmonyAIR E-Series Surgical Lighting Systems. Recalled by Steris Corporation. Units affected: 59 units.
Why was this product recalled?
The DFPM yoke assembly ("yoke assembly") may detach from the spring arm. Should the yoke assembly detach from the spring arm, the yoke assembly would not be expected to fully detach but rather hang from the wiring within the yoke assembly.
Which agency issued this recall?
This recall was issued by the FDA Devices on August 6, 2025. Severity: Moderate. Recall number: Z-2178-2025.