FDA Devices Moderate Class II Terminated

Dolphin Inflation Device, REF 0185NA. Rx only, Sterilized using Ethylene Oxide. DOLPHIN is a single-use, sterile inflation device used in cardiovascular procedures to pressurize (inflate) and depressurize (deflate) balloon catheters. The device itself consists of two components - the inflation device and large volume syringe (30 cc). The manually operation of the DOLPHIN is achieved by the manipulation of a large handle to drive a piston housed within the body of the device. Careful a

Reported: June 7, 2017 Initiated: April 12, 2017 #Z-2180-2017

Product Description

Dolphin Inflation Device, REF 0185NA. Rx only, Sterilized using Ethylene Oxide. DOLPHIN is a single-use, sterile inflation device used in cardiovascular procedures to pressurize (inflate) and depressurize (deflate) balloon catheters. The device itself consists of two components - the inflation device and large volume syringe (30 cc). The manually operation of the DOLPHIN is achieved by the manipulation of a large handle to drive a piston housed within the body of the device. Careful and controlled inflation is achieved by rotating the handle clockwise. During inflation a unique cam locking mechanism maintains pressure even if the user lets go of the device. Instantaneous deflation, regardless of balloon size, is made possible by the release of the dual locks located on device as well as a large, 30 cc syringe that is sold with each Dolphin Inflation Device. All the while during inflation or deflation pressure is displayed and can be monitored on a large analog gauge mounted on top of the device.

Reason for Recall

The manufacturer of the DOLPHIN inflation device, Perouse Medical, has initiated a recall of the device due to a potential defect in the sterile barrier. Use of affected product could pose a risk of microbial contamination, leading to infection.

Details

Recalling Firm
Vascular Solutions, Inc.
Units Affected
4,110
Distribution
US Distribution including PR and to the states of : LA, TN, CA, SC, OR, WA, OK, NV, HI, OH, TX.
Location
Maple Grove, MN

Frequently Asked Questions

What product was recalled?
Dolphin Inflation Device, REF 0185NA. Rx only, Sterilized using Ethylene Oxide. DOLPHIN is a single-use, sterile inflation device used in cardiovascular procedures to pressurize (inflate) and depressurize (deflate) balloon catheters. The device itself consists of two components - the inflation device and large volume syringe (30 cc). The manually operation of the DOLPHIN is achieved by the manipulation of a large handle to drive a piston housed within the body of the device. Careful and controlled inflation is achieved by rotating the handle clockwise. During inflation a unique cam locking mechanism maintains pressure even if the user lets go of the device. Instantaneous deflation, regardless of balloon size, is made possible by the release of the dual locks located on device as well as a large, 30 cc syringe that is sold with each Dolphin Inflation Device. All the while during inflation or deflation pressure is displayed and can be monitored on a large analog gauge mounted on top of the device.. Recalled by Vascular Solutions, Inc.. Units affected: 4,110.
Why was this product recalled?
The manufacturer of the DOLPHIN inflation device, Perouse Medical, has initiated a recall of the device due to a potential defect in the sterile barrier. Use of affected product could pose a risk of microbial contamination, leading to infection.
Which agency issued this recall?
This recall was issued by the FDA Devices on June 7, 2017. Severity: Moderate. Recall number: Z-2180-2017.