Severity
Moderate
FDA Devices recall · Reported August 15, 2012
Siemens Healthcare Diagnostics has confirmed that partial insertion of the power cord on the Dimension Reagent Management System (RMS) power input module can cause the plug to ove…
Siemens Healthcare Diagnostics, Inc. recalled Dimension(R) Reagent Management System (RMS) - Power Cord. The Siemens Healthcare Diag… - a moderate-severity action.
Dimension(R) Reagent Management System (RMS) - Power Cord. The Siemens Healthcare Diag… was recalled by Siemens Healthcare Diagnostics, Inc. in August 15, 2012. Reason: Siemens Healthcare Diagnostics has confirmed that partial insertion of the power cord on the Dimension Reagen…. Check the official notice for the remedy. Verify recall #Z-2182-2012 with the FDA Devices before acting.
The recall
Siemens Healthcare Diagnostics, Inc. issued this moderate-severity FDA Devices recall-Siemens Healthcare Diagnostics has confirmed that partial insertion of the power cord on the Dimension Reagen….
Sourced from official FDA Devices enforcement records. Verify recall #Z-2182-2012 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-2182-2012) was formally reported on August 15, 2012, with the manufacturer initiating the action on July 17, 2012. It is classified under Moderate severity (Class II), with a current status of Terminated. Siemens Healthcare Diagnostics, Inc. is listed as the recalling firm, operating out of Newark, DE. Federal records list the affected scope as 2,554.
The documented reason for this recall is: Siemens Healthcare Diagnostics has confirmed that partial insertion of the power cord on the Dimension Reagent Management System (RMS) power input module can cause the plug to overheat and potentially create a fire haza… Distribution data in the federal record shows the product reached: Distributed in the US including Washington, DC, Puerto Rico and the states of AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH,…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
2,554
Related Recalls
6
3 from same agency
Dimension(R) Reagent Management System (RMS) - Power Cord. The Siemens Healthcare Diagnostics Dimension(R) clinical chemistry analyzers are floor model, fully automated, microprocessor-controlled, integrated instrument systems that use prepackaged Flex(R) reagent test cartridges to measure a variety of analytes in human body fluids.
Siemens Healthcare Diagnostics has confirmed that partial insertion of the power cord on the Dimension Reagent Management System (RMS) power input module can cause the plug to overheat and potentially create a fire hazard. RMS without a power cord retention clip may be affected.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-2182-2012 |
| Date reported | August 15, 2012 |
| Date initiated | July 17, 2012 |
| Recalling firm | Siemens Healthcare Diagnostics, Inc. |
| Firm location | Newark, DE |
| Affected scope | 2,554 |
| Distribution | Distributed in the US including Washington, DC, Puerto Rico and the states of AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, SC, SD,… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported August 15, 2012. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.