Severity
Moderate
FDA Devices recall · Reported July 27, 2016
During a procedure the Peak Skin Dose (PSD) value was displayed on the DTS larger than the DTS expected value. It was found that the software incorrectly calculated the PSD value …
Toshiba American Medical Systems Inc recalled Toshiba America Medical Systems Angio WorkStation: XIDF-AWS801v6.00, v6.01, and V6.10 Sys… - a moderate-severity action.
Toshiba America Medical Systems Angio WorkStation: XIDF-AWS801v6.00, v6.01, and V6.10 Sys… was recalled by Toshiba American Medical Systems Inc in July 27, 2016. Reason: During a procedure the Peak Skin Dose (PSD) value was displayed on the DTS larger than the DTS expected value…. Check the official notice for the remedy. Verify recall #Z-2188-2016 with the FDA Devices before acting.
The recall
Toshiba American Medical Systems Inc issued this moderate-severity FDA Devices recall-During a procedure the Peak Skin Dose (PSD) value was displayed on the DTS larger than the DTS expected value….
Sourced from official FDA Devices enforcement records. Verify recall #Z-2188-2016 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-2188-2016) was formally reported on July 27, 2016, with the manufacturer initiating the action on February 24, 2016. It is classified under Moderate severity (Class II), with a current status of Terminated. Toshiba American Medical Systems Inc is listed as the recalling firm, operating out of Tustin, CA. Federal records list the affected scope as 23.
The documented reason for this recall is: During a procedure the Peak Skin Dose (PSD) value was displayed on the DTS larger than the DTS expected value. It was found that the software incorrectly calculated the PSD value when the x-ray condition of the frontal … Distribution data in the federal record shows the product reached: US Distribution to the states of : NC, NY, TX, GA, CA, AZ, IL, FL,MA, MO and DE.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
23
Related Recalls
6
3 from same agency
Toshiba America Medical Systems Angio WorkStation: XIDF-AWS801v6.00, v6.01, and V6.10 System: INFX-8000V Bi-Plane System
During a procedure the Peak Skin Dose (PSD) value was displayed on the DTS larger than the DTS expected value. It was found that the software incorrectly calculated the PSD value when the x-ray condition of the frontal plane and the lateral plane are the same. It incorrectly used in the calculation a dose value from a previous exposure.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-2188-2016 |
| Date reported | July 27, 2016 |
| Date initiated | February 24, 2016 |
| Recalling firm | Toshiba American Medical Systems Inc |
| Firm location | Tustin, CA |
| Affected scope | 23 |
| Distribution | US Distribution to the states of : NC, NY, TX, GA, CA, AZ, IL, FL,MA, MO and DE. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported July 27, 2016. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.